0 768

Cited 10 times in

Removal of a broken trigen intertan intertrochanteric antegrade nail

Authors
 Xuan-Lin Zheng  ;  Young-Chang Park  ;  Sungmin Kim  ;  Haemosu An  ;  Kyu-Hyun Yang 
Citation
 INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, Vol.48(2) : 557-559, 2017 
Journal Title
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED
ISSN
 0020-1383 
Issue Date
2017
MeSH
Biomechanical Phenomena ; Bone Nails* ; Device Removal/methods* ; Equipment Failure ; Foreign-Body Migration*/diagnostic imaging ; Foreign-Body Migration*/surgery ; Fracture Fixation, Intramedullary/adverse effects ; Fracture Fixation, Intramedullary/instrumentation* ; Hip Fractures/diagnostic imaging ; Hip Fractures/surgery* ; Humans ; Joint Instability/diagnostic imaging ; Joint Instability/surgery* ; Male ; Middle Aged ; Radiography* ; Reoperation
Keywords
Lag screw ; Mechanical failure ; Nail ; Trigen intertan
Abstract
Implant breakage is a serious complication after cephalomedullary nailing for unstable intertrochanteric fracture. Failure usually occurs at the lag screw hole in the nail body. On the other hand, lag screw failure is extremely rare and occurs around the nail-lag screw junction. We experienced rare mechanical failure of the Intertan nail, which showed breakage at the lag screw hole and failure of the integrated compression screw underneath the main lag screw.
Full Text
http://www.sciencedirect.com/science/article/pii/S0020138316308208
DOI
10.1016/j.injury.2016.12.014
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Orthopedic Surgery (정형외과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Sung Min(김성민)
Park, Young Chang(박영창) ORCID logo https://orcid.org/0000-0003-3726-4707
An, Haemosu(안해모수)
Yang, Kyu Hyun(양규현) ORCID logo https://orcid.org/0000-0001-7183-588X
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/154299
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links