Can FLAIR hyperintense vessel (FHV) signs be influenced by varying MR parameters and flow velocities?

Other Titles
 MR parameter와 혈류 속도의 변화가 FLAIR hyperintense vessel (FHV) sign에 미치는 영향 
Authors
 안성준 
Department
 Dept. of Radiology (영상의학교실) 
Issue Date
2015
Description
의과대학/박사
Abstract
Purpose: Fluid-attenuated inversion recovery (FLAIR) hyperintense vessels (FHVs) have been used to assess leptomeningeal collateral flow in acute ischemic stroke. However, prior FHVs studies showed inconsistent results, which may be ascribable to different MR parameters used. The aim of our study was to evaluate whether FHVs could be influenced by varying MR parameters and flow velocities, using a flow phantom.



Material and Methods: A total of 512 sets of FLAIRs were performed with varying parameters and flow velocities, using a flow phantom. Flow phantom was manufactured with 3.5 % agarose solution, an 8-mm-inner diameter silicone tube and non-pulsatile pump. Varying MR parameters were repetition time (TR)/inversion time (TI), echo time (TE), flip angle (FA) of refocusing pulse and periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER). The signal intensity of flow were measured and regarded as the degree of FHVs. Simple and multiple linear regression analyses were applied to evaluate the association between the degree of FHVs and varying MR parameters as well as flow velocities.



Results: On univariate analysis, PROPELLER technique (R2 = 0.448) demonstrated strongest correlation with the degree of FHV, followed by flow velocities (R2 = 0.204), FA (R2 = 0.126) and TE (R2 = 0.031), whereas TR/TI showed no significant correlations. On multivariate analysis, TE, FA, PROPELLER technique and flow velocities were independent factors influencing the degree of FHVs (<0.001).



Conclusion: Flow velocities, FA of refocusing pulse, TE and PROPELLER technique significantly affected the degree of FHVs. Optimized MR parameters should be used consistently in future studies, which may provide more reliable results.
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 3. Dissertation
Yonsei Authors
Ahn, Sung Jun(안성준) ORCID logo https://orcid.org/0000-0003-0075-2432
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/148692
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links