0 333

Cited 0 times in

The Application of Bioelectrical Impedance to Monitor Leg Movement

Authors
 Chul-Gyu SONG  ;  Deok Won KIM 
Citation
 IEICE TRANSACTIONS ON INFORMATION AND SYSTEMS, Vol.E88-D(1) : 164-168, 2005 
Journal Title
IEICE TRANSACTIONS ON INFORMATION AND SYSTEMS
ISSN
 0916-8532 
Issue Date
2005
MeSH
electrical bio-impedance ; joint movement ; human motion ; impedance method
Keywords
electrical bio-impedance ; joint movement ; human motion ; impedance method
Abstract
The purpose of this study is to provide a new approach for detection using bio-impedance. This impedance is measured by the four-electrode method. As the impedance changes resulting from ankle, knee, and hip movements depended heavily on electrode placement, we determined the optimal electrode configurations for those movements by searching for high correlation coefficients, large impedance changes, and minimum interferences in ten subjects (Age: 204). Our optimal electrode configurations showed very strong relationships between the ankle joint angle and ankle impedance (γ = -0.9130.03), between the knee joint angle and knee impedance (γ = 0.9440.02), and between the hip joint angle and hip impedance (γ = 0.8230.08). This study showed the possibility that lower leg movement could be easily measured by impedance measurement system with two pairs of skin-electrodes.
Full Text
http://search.ieice.org/bin/summary.php?id=e88-d_1_164&category=D&year=2005&lang=E&abst=
DOI
10.1093/ietisy/E88-D.1.164
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Medical Engineering (의학공학교실) > 1. Journal Papers
Yonsei Authors
Kim, Deok Won(김덕원)