515 579

Cited 0 times in

혈액 투석관 내 혈류 측정을 위한 생체외 실험모델

Other Titles
 Flow Measurement in the Vascular Access Conduit: an Ex Vivo Experimental Model 
Authors
 이종훈  ;  허규하  ;  김유선  ;  조희은  ;  박기일 
Citation
 JOURNAL OF THE KOREAN SURGICAL SOCIETY , Vol.68(4) : 253-257, 2005 
Journal Title
JOURNAL OF THE KOREAN SURGICAL SOCIETY (대한외과학회지)
ISSN
 1226-0053 
Issue Date
2005
MeSH
Pressure measurement ; Flow rate ; Ex vivo experiment
Keywords
Pressure measurement ; Flow rate ; Ex vivo experiment
Abstract
Purpose: Previously, we developed a new method to calculate the flow rate in the hemodialysis vascular conduit based on Bernoulli's theory for surveillance of the arteriovenous fistula (AVF) function. However, the calculated flow rate would be different from the true flow rate because due to various factors. To compare the true flow rate, with intra- conduit pressure, and the calculated flow rate, an ex vivo experimental model was developed.
Methods: The arterial end of the vascular conduit was connected to a saline-filled bottle, with the venous end connected to a flow meter to control the flow rate. By monitoring the change in the true flow rate (Q) with the flow meter, each arterial and venous static pressure (pA, pV) and total pressure (pT) were observed. Using these parameters, the intra-conduit flow rates (QA, QV) were calculated by Bernoulli's equation. Finally, we compared the pA or pV with Q, and calculated the difference between the QA or QV and Q.
Results: There were no statistical differences between any of the pressure measurement during the 5 consecutive 5 experiments (P<0.05). The static pressure (pA or pV) was closely correlated with Q (pA, R2=0.950, P=0.000; pV, R2= 0.952, P=0.000). The calculated flow rate (QA or QV) was not in complete in accord with Q, but was closely correlated (QA, R2=0.961, P=0.000; QV, R2=0.961, P=0.000).
Conclusion: The pressure measurement and calculated flow rate indicate the nature of the true flow rate in the vascular conduit.
Files in This Item:
T200500161.pdf Download
DOI
OAK-2005-02778
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Surgery (외과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Yu Seun(김유선) ORCID logo https://orcid.org/0000-0002-5105-1567
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/147419
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links