0 706

Cited 12 times in

Usefulness of a lead shielding device for reducing the radiation dose to tissues outside the primary beams during CT

DC Field Value Language
dc.contributor.author김주희-
dc.contributor.author유정식-
dc.contributor.author정재준-
dc.contributor.author조은석-
dc.date.accessioned2015-12-28T10:58:03Z-
dc.date.available2015-12-28T10:58:03Z-
dc.date.issued2014-
dc.identifier.issn0033-8362-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/138408-
dc.description.abstractPURPOSE: This study was done to investigate the efficacy of a lead shield in protecting the tissues outside the primary beams, such as the breast and thyroid, by measurement of the entrance skin dose during CT of the brain, neck, abdomen, and lumbar spine. MATERIALS AND METHODS: Institutional Review Board approval was obtained. This study included 150 patients (male:female 25:125, age range 15-45 years). In females, brain, lumbar spine, and abdominal CT scans, pre-/post-contrast neck CT scans, and post-contrast liver dynamic CT scans were performed. In males, brain CT scans only were performed. Breast shielding was performed in all females, and thyroid shielding was conducted in patients with brain CT. During all CT studies, the left breast or left thyroid was shielded using a lead shield, and the contralateral side was left unshielded. Thus, each breast or thyroid measurement had its own control for the same demographic data. The efficacy of the shielding of both breasts and thyroids during CT was assessed. RESULTS: During brain, abdominal, lumbar, pre-/post-contrast neck, and post-contrast liver dynamic CT, 33.5, 26.0, 17.4, 26.5, and 16.2 % of the breast skin dose were reduced, respectively. During brain CT, the thyroid skin dose was reduced by 17.9 % (females) and 20.6 % (males). There were statistically significant differences in the skin doses of shielded organs (p < 0.05). Breast shielding during neck and liver dynamic CT was the most effective compared with breast or thyroid shielding during other CT scans. CONCLUSIONS: We recommend breast shielding during neck and liver dynamic CT in young female patients to avoid unnecessary radiation exposure.-
dc.description.statementOfResponsibilityopen-
dc.relation.isPartOfRADIOLOGIA MEDICA-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.subject.MESHAdolescent-
dc.subject.MESHAdult-
dc.subject.MESHBreast/radiation effects-
dc.subject.MESHFemale-
dc.subject.MESHHumans-
dc.subject.MESHLead*-
dc.subject.MESHMale-
dc.subject.MESHMiddle Aged-
dc.subject.MESHRadiation Dosage*-
dc.subject.MESHRadiation Protection/methods-
dc.subject.MESHRadiation-Protective Agents*-
dc.subject.MESHSkin/radiation effects-
dc.subject.MESHThyroid Gland/radiation effects-
dc.subject.MESHTomography, X-Ray Computed*-
dc.subject.MESHYoung Adult-
dc.titleUsefulness of a lead shielding device for reducing the radiation dose to tissues outside the primary beams during CT-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Radiology (영상의학)-
dc.contributor.googleauthorJae-Joon Chung-
dc.contributor.googleauthorEun-Suk Cho-
dc.contributor.googleauthorSung Min Kang-
dc.contributor.googleauthorJeong-Sik Yu-
dc.contributor.googleauthorDae Jung Kim , Joo Hee Kim-
dc.identifier.doi10.1007/s11547-014-0421-0-
dc.admin.authorfalse-
dc.admin.mappingfalse-
dc.contributor.localIdA00951-
dc.contributor.localIdA02500-
dc.contributor.localIdA03712-
dc.contributor.localIdA03881-
dc.relation.journalcodeJ02594-
dc.identifier.eissn1826-6983-
dc.identifier.pmid24903707-
dc.identifier.urlhttp://link.springer.com/article/10.1007%2Fs11547-014-0421-0-
dc.contributor.alternativeNameKim, Joo Hee-
dc.contributor.alternativeNameYu, Jeong Sik-
dc.contributor.alternativeNameChung, Jae Joon-
dc.contributor.alternativeNameCho, Eun Suk-
dc.contributor.affiliatedAuthorKim, Joo Hee-
dc.contributor.affiliatedAuthorYu, Jeong Sik-
dc.contributor.affiliatedAuthorChung, Jae Joon-
dc.contributor.affiliatedAuthorCho, Eun Suk-
dc.rights.accessRightsfree-
dc.citation.titleLa Radiologia Medica-
dc.citation.volume119-
dc.citation.number12-
dc.citation.number12-
dc.citation.startPage951-
dc.citation.endPage957-
dc.identifier.bibliographicCitationRADIOLOGIA MEDICA, Vol.119(12) : 951-957, 2014-
dc.identifier.rimsid49158-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.