두개골 팬텀을 이용한 다검출기 CT 3차원 영상에서의 거리측정을 통한 정량적 영상특성 평가

Other Titles
 Quantitative Evaluation of the Accuracy of 3D Imaging with Multi-Detector Computed Tomography Using Human Skull Phantom 
Authors
 김동욱  ;  정해조  ;  김희중  ;  김기덕  ;  유영일  ;  김새롬 
Citation
 Korean Journal of Medical Physics (의학물리), Vol.14(2) : 131-140, 2003 
Journal Title
Korean Journal of Medical Physics(의학물리)
ISSN
 1226-5829 
Issue Date
2003
MeSH
multi-detector computed tomography ; volume rendering ; 3-D reconstruction ; slice thick-ness ; Quantitative analysis
Keywords
multi-detector computed tomography ; volume rendering ; 3-D reconstruction ; slice thick-ness ; Quantitative analysis
Abstract
As the importance of accuracy in measurings of 3-D anatomical structures continues to be stressed, an objective and quantitative of assessing image quality and accuracy of 3-D volume-rendered images is required. The purpose of this study was to evaluate the quantitative accuracy of 3-D rendered images obtained with MDCT, scanned at various scanning parameters (scan modes, slice thicknesses and reconstruction slice thickness). Twelve clinically significant points that play an important role for the craniofacial bone in plastic surgery and dentistry were marked on the surface of a dry human skull. The direct distances between the reference points were defined as gold standards to assess the measuring errors of 3-D images. Then, we scanned the specimen with acquisition parameters of 300 mA, In kVp, and 1.0 sec scan time in axial and helical scan modes (pitch 3:1 and 6:1) at 1,25 mm, 2.50 mm, 3.75 mm and 5.00 mm slice thicknesses. We performed 3-D visualizations and distance measurements with volumetric analysis software and statistically evaluated the quantitative accuracy of distance measurements. The accuracy of distance measurements on the 3-D images acquired with 1.25, 2.50, 3,75 and 5.00 mm slice thickness were 48%, 33%, 23%, 14%, respectively, and those of the reconstructed 1.25 mm were 53%, 41%, 43%, 36% respectively. Meanwhile, there were insignificant statistical differences (P-value<0.05) in the accuracy of the distance measurements of 3-D images reconstructed with 1.25 mm thickness. In conclusion, slice thickness, rather than scan mode, influenced the quantitative accuracy of distance measurements in 3-D rendered images with MDCT. The quantitative analysis of distance measurements may be a useful tool for evaluating the accuracy of 3-D rendered images used in diagnosis, surgical planning, and radiotherapeutic treatment.
Files in This Item:
T200305240.pdf Download
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Advanced General Dentistry (통합치의학과) > 1. Journal Papers
Yonsei Authors
Kim, Kee Deog(김기덕) ORCID logo https://orcid.org/0000-0003-3055-5130
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/114049
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links