베르누이 정리를 이용한 혈액 투석관내 혈류량 측정법 개발

Other Titles
 Development of Simple Flow Measurement in the Vascular Access Conduit Based on the Bernoulli's Theory 
Authors
 이종훈  ;  송순영  ;  박기일  ;  김유선  ;  김순일  ;  허규하  ;  양성자  ;  윤수영  ;  조현경 
Citation
 Journal of the Korean Society for Vascular Surgery (대한혈관외과학회지), Vol.20(1) : 115-124, 2004 
Journal Title
Journal of the Korean Society for Vascular Surgery(대한혈관외과학회지)
ISSN
 2233-9779 
Issue Date
2004
Keywords
Hemodialysis ; Ateriovenous fistula ; Flow rates ; 혈액 투석 ; 내동정맥루 ; 혈류량
Abstract
Purpose: Early detection of arteriovenous fistula (AVF) dysfunction in hemodialysis patients and prompt corrective procedures reduces the AVF thrombosis rates and lengthens access survival. We tried to develop a new simple and cheap bedside measurement technique based on the Bernoulli´s theory.
Method: From a total of 20 case of vascular accesses for hemodialysis, of at least 3 months of construction, we twicely measured the AVF flow rate (QD) with Doppler ultrasonography and vascular conduit pressure. Four kinds of pressure were measured: tubing set free from dialysis machine and positioned on the patient´s bed (PrF), two kinds of artificial stenosis made with tourniquet (PrS1, PrS2), pump flow rate at 100ml/min (Pr100), and pump off (Pr0). We calculated the flow rate of vascular conduit (QF) with PrF and mean arterial pressure on Bernouli´s equation, and QF was compared with QD.
Result: AVF was 26.0⁑28.6 (3∼108) months after operation, with five cases (including 2 PTFE grafts) using brachial artery. PrF was closely correlated with Pr100 (R2=0.914), and inversely correlated with QD (R2=-0.026). QF was poorly correlated with QD (R2=0.003). There was no statistical difference in the double pressure measurement (P>0.05), but there was differenence in QD (P<0.05). When artificial stenosis was made, the pressures increased, and the calculated flow rates decreased in every patients. Thrombosis or stenosis was detected in all patients with decrement of QF, but not in all with decrement of QD.
Conclusion: Pressure measurement and calculated flow rate in dialysis vascular conduit represent alterations of AVF flow rate. However its value in long-term follow up awaits further study with accurate constant number.
Files in This Item:
T200400222.pdf Download
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Surgery (외과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Soon Il(김순일) ORCID logo https://orcid.org/0000-0002-0783-7538
Kim, Yu Seun(김유선) ORCID logo https://orcid.org/0000-0002-5105-1567
Huh, Kyu Ha(허규하) ORCID logo https://orcid.org/0000-0003-1364-6989
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/111980
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links