Reduction of B1 sensitivity in selective single-slab 3D turbo spin echo imaging with very long echo trains

DC Field Value Language
dc.contributor.author박재석-
dc.date.accessioned2015-04-24T16:54:31Z-
dc.date.available2015-04-24T16:54:31Z-
dc.date.issued2009-
dc.identifier.issn0740-3194-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/104399-
dc.description.abstractSingle-slab 3D turbo/fast spin echo (SE) imaging with very long echo trains was recently introduced with slab selection using a highly selective excitation pulse and short, nonselective refocusing pulses with variable flip angles for high imaging efficiency. This technique, however, is vulnerable to image degradation in the presence of spatially varying B(1) amplitudes. In this work we develop a B(1) inhomogeneity-reduced version of single-slab 3D turbo/fast SE imaging based on the hypothesis that it is critical to achieve spatially uniform excitation. Slab selection was performed using composite adiabatic selective excitation wherein magnetization is tipped into the transverse plane by a nonselective adiabatic-half-passage pulse and then slab is selected by a pair of selective adiabatic-full-passage pulses. Simulations and experiments were performed to evaluate the proposed technique and demonstrated that this approach is a simple and efficient way to reduce B(1) sensitivity in single-slab 3D turbo/fast SE imaging with very long echo trains-
dc.description.statementOfResponsibilityopen-
dc.format.extent1060~1066-
dc.relation.isPartOfMAGNETIC RESONANCE IN MEDICINE-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.subject.MESHAlgorithms*-
dc.subject.MESHEcho-Planar Imaging/methods*-
dc.subject.MESHHumans-
dc.subject.MESHImage Enhancement/methods*-
dc.subject.MESHImage Interpretation, Computer-Assisted/methods*-
dc.subject.MESHImaging, Three-Dimensional/methods*-
dc.subject.MESHReproducibility of Results-
dc.subject.MESHSensitivity and Specificity-
dc.subject.MESHSpin Labels-
dc.subject.MESHWhole Body Imaging/methods*-
dc.titleReduction of B1 sensitivity in selective single-slab 3D turbo spin echo imaging with very long echo trains-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Radiology (영상의학)-
dc.contributor.googleauthorJaeseok Park-
dc.contributor.googleauthorJohn P. Mugler III-
dc.contributor.googleauthorTimothy Hughes-
dc.identifier.doi10.1002/mrm.22070-
dc.admin.authorfalse-
dc.admin.mappingfalse-
dc.contributor.localIdA01635-
dc.relation.journalcodeJ02179-
dc.identifier.eissn1522-2594-
dc.identifier.pmid19585600-
dc.identifier.urlhttp://onlinelibrary.wiley.com/doi/10.1002/mrm.22070/abstract-
dc.subject.keywordmagnetic resonance imaging (MRI)-
dc.subject.keywordspin echo-
dc.subject.keywordthree‐dimensional-
dc.subject.keywordvariable flip angle-
dc.subject.keywordB 1 sensitivity-
dc.subject.keywordadiabatic excitation-
dc.contributor.alternativeNamePark, Jae Seok-
dc.contributor.affiliatedAuthorPark, Jae Seok-
dc.citation.volume62-
dc.citation.number4-
dc.citation.startPage1060-
dc.citation.endPage1066-
dc.identifier.bibliographicCitationMAGNETIC RESONANCE IN MEDICINE, Vol.62(4) : 1060-1066, 2009-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiology (영상의학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.