231 508

Cited 2 times in

Electro-Mechanical Safety Testing of Portable ECG Devices for Home Healthcare Usage

Authors
 Ju Hyun Kim  ;  Eun Jeong Cho  ;  Chung Ki Lee  ;  Su Kang Park  ;  Byung Chae Lee  ;  Sun Kook Yoo 
Citation
 HEALTHCARE INFORMATICS RESEARCH, Vol.16(1) : 30-35, 2010 
Journal Title
HEALTHCARE INFORMATICS RESEARCH
ISSN
 2093-3681 
Issue Date
2010
Keywords
Electro Medical Device ; Electrochemical Techniques ; Equipment Safety ; Home-Healthcare Device ; Standards
Abstract
OBJECTIVES: This paper suggests the experimental guidelines to evaluate the electro-mechanical safety of belt type equipment. The electro-mechanical safety was determined by using the International Electrotechnical Commission guidelines, which are widely used as important factors for assessing the electro-mechanical safety of belt type equipment. However, the local guidelines on wearable healthcare sensors are currently not well-established. Therefore, safety guidelines suited for the actual circumstances in Korea are required, and this paper attempts to try a new experimental safety test procedure of the wearable healthcare sensor. METHODS: This belt type device measures the electrocardiogram (ECG) and heart rates by attaching to the chest. Examination lists were selected by analyzing the common standards ofelectro-mechanical safety (IEC 60601-1) and environment tests (IEC 60068-1, IEC 60068-2) of home-healthcare equipment. RESULTS: The essential electrical safety, which was required for the RS300G3 as a medical device, was evaluated, and most of the examination lists were selected by considering the circumstances of the users. The device passed all the selected examinable lists that are applicable to the Korean environment. CONCLUSIONS: This study has limitations to estimate and to conduct electro-mechanical safety experiments because our study focused on the belt type of heart-rates equipment. We are not taking into account the overall electro-mechanical home-healthcare measurements. According to industrial and technological development, there are infinite possibilities for the advancement of home-healthcare equipment, so more examination lists for safety are being added in addition to what we have done
Files in This Item:
T201003448.pdf Download
DOI
10.4258/hir.2010.16.1.30
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Medical Engineering (의학공학교실) > 1. Journal Papers
Yonsei Authors
Kim, Ju Hyun(김주현)
Yoo, Sun Kook(유선국) ORCID logo https://orcid.org/0000-0002-6032-4686