439 650

Cited 39 times in

A Pilot Study of Sequential Capsule Endoscopy Using MiroCam and PillCam SB Devices with Different Transmission Technologies

Authors
 Hee Man Kim  ;  Yoon Jae Kim  ;  Hong Jeong Kim  ;  Semi Park  ;  Jeong Youp Park  ;  Sung Kwan Shin  ;  Jae Hee Cheon  ;  Sang Kil Lee  ;  Yong Chan Lee  ;  Seung Woo Park  ;  Seungmin Bang  ;  Si Young Song 
Citation
 GUT AND LIVER, Vol.4(2) : 192-200, 2010 
Journal Title
GUT AND LIVER
ISSN
 1976-2283 
Issue Date
2010
Keywords
Capsule endoscopy ; Diagnosis ; Feasibility study ; Gastrointestinal hemorrhage
Abstract
BACKGROUND/AIMS: Studies have investigated the use of different types of radiofrequency capsules for comparison or sequential capsule endoscopy, but none have compared the MiroCam device - which utilizes a novel data transmission technology - with other capsules. This study compared the feasibility of sequential capsule endoscopy using the MiroCam and PillCam SB devices, which employ different transmission technologies.

METHODS: Patients with diseases requiring capsule endoscopy were enrolled. After a 12-hour fast, one randomly selected capsule was swallowed. The second capsule was swallowed once fluoroscopy had indicated that the first capsule had migrated below the gastric outlet.

RESULTS: The total operating time in 24 patients was 702+/-60 min (mean+/-SD) for the MiroCam and 446+/-28 min for the PillCam SB (p<0.0001). The rate of a complete examination to the cecum was 83.3% for the MiroCam and 58.3% for the PillCam SB (p=0.031). Diagnostic yields for the MiroCam, PillCam SB, and sequential capsule endoscopy were 45.8%, 41.7%, and 50.0%, respectively. The agreement rate between the two capsules was 87.5%, with a kappa value of 0.74. Electrical interference in data transmission between the two capsules was not observed, but temporary visual interferences were observed in seven patients (29.2%).

CONCLUSIONS: Sequential capsule endoscopy with the MiroCam and PillCam SB produced slight but nonsignificant increases in the diagnostic yield, and the two capsules did not exhibit electrical interference. A larger trial is necessary for elucidating the usefulness of sequential capsule endoscopy.
Files in This Item:
T201001834.pdf Download
DOI
10.5009/gnl.2010.4.2.192
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers
Yonsei Authors
Park, Seung Woo(박승우) ORCID logo https://orcid.org/0000-0001-8230-964X
Park, Jeong Youp(박정엽) ORCID logo https://orcid.org/0000-0003-0110-8606
Bang, Seungmin(방승민) ORCID logo https://orcid.org/0000-0001-5209-8351
Song, Si Young(송시영) ORCID logo https://orcid.org/0000-0002-1417-4314
Shin, Sung Kwan(신성관) ORCID logo https://orcid.org/0000-0001-5466-1400
Lee, Sang Kil(이상길) ORCID logo https://orcid.org/0000-0002-0721-0364
Lee, Yong Chan(이용찬) ORCID logo https://orcid.org/0000-0001-8800-6906
Cheon, Jae Hee(천재희) ORCID logo https://orcid.org/0000-0002-2282-8904
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/101209
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links