Cited 1 times in

The instability of commercial control materials in quality control of mean corpuscular volume

DC Field Value Language
dc.contributor.author송예진-
dc.contributor.author송재우-
dc.date.accessioned2015-01-06T17:01:32Z-
dc.date.available2015-01-06T17:01:32Z-
dc.date.issued2014-
dc.identifier.issn0009-8981-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/99224-
dc.description.abstractBACKGROUND: Mean corpuscular volume (MCV) of stabilized whole blood used for quality control (QC) of hematology analyzers exhibits a tendency to increase during storage. The aim of this study is to evaluate the extent of biases over time with 3 most widely used control materials and to map out a strategy to overcome the data shift of MCV on daily QC practice. METHODS:QC results of TESTPoint tested by ADVIA 2120i, e-CHEK tested by XE 2100, and 6C Cell Control tested by DxH 800 were analyzed. RESULTS:MCV of all control materials showed a tendency to increase over time. By the fifth week, most of the materials showed biases larger than one standard deviation, with some exceeding a bias of four standard deviations. CONCLUSIONS:Laboratories should apply appropriate QC strategies in MCV tests by considering their individual quality and efficiency requirements.-
dc.description.statementOfResponsibilityopen-
dc.format.extent11~15-
dc.relation.isPartOfCLINICA CHIMICA ACTA-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.subject.MESHAutomation-
dc.subject.MESHBlood Preservation*-
dc.subject.MESHErythrocyte Indices*-
dc.subject.MESHFalse Negative Reactions-
dc.subject.MESHFalse Positive Reactions-
dc.subject.MESHHumans-
dc.subject.MESHQuality Control-
dc.subject.MESHReproducibility of Results-
dc.subject.MESHSensitivity and Specificity-
dc.subject.MESHSpecimen Handling-
dc.titleThe instability of commercial control materials in quality control of mean corpuscular volume-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentYonsei Biomedical Research Center (연세의생명연구원)-
dc.contributor.googleauthorSue Jung Kim-
dc.contributor.googleauthorEun Young Lee-
dc.contributor.googleauthorYea Jin Song-
dc.contributor.googleauthorJaewoo Song-
dc.identifier.doi10.1016/j.cca.2014.04.004-
dc.admin.authorfalse-
dc.admin.mappingfalse-
dc.contributor.localIdA02041-
dc.contributor.localIdA02054-
dc.relation.journalcodeJ00543-
dc.identifier.eissn1873-3492-
dc.identifier.pmid24769249-
dc.identifier.urlhttp://www.sciencedirect.com/science/article/pii/S0009898114001594-
dc.subject.keywordAutomated hematology analyzer-
dc.subject.keywordErythrocyte indices-
dc.subject.keywordMean corpuscular volume-
dc.subject.keywordQuality control-
dc.subject.keywordStabilized blood-
dc.contributor.alternativeNameSong, Yea Jin-
dc.contributor.alternativeNameSong, Jae Woo-
dc.contributor.affiliatedAuthorSong, Yea Jin-
dc.contributor.affiliatedAuthorSong, Jae Woo-
dc.rights.accessRightsfree-
dc.citation.volume434-
dc.citation.startPage11-
dc.citation.endPage15-
dc.identifier.bibliographicCitationCLINICA CHIMICA ACTA, Vol.434 : 11-15, 2014-
dc.identifier.rimsid55945-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Yonsei Biomedical Research Center (연세의생명연구원) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.