1 579

Cited 43 times in

The ability of stroke volume variation measured by a noninvasive cardiac output monitor to predict fluid responsiveness in mechanically ventilated children

Authors
 Ji Yeon Lee  ;  Ji Young Kim  ;  Chang Hyu Choi  ;  Hong Soon Kim  ;  Kyung Cheon Lee  ;  Hyun Jeong Kwak 
Citation
 PEDIATRIC CARDIOLOGY, Vol.35(2) : 289-294, 2014 
Journal Title
PEDIATRIC CARDIOLOGY
ISSN
 0172-0643 
Issue Date
2014
MeSH
Cardiac Surgical Procedures ; Child ; Child, Preschool ; Circadian Rhythm/physiology* ; Echocardiography/methods* ; Female ; Fluid Therapy/methods* ; Follow-Up Studies ; Heart Septal Defects, Ventricular/surgery ; Humans ; Infant ; Male ; Monitoring, Physiologic/methods* ; Postoperative Care/methods* ; Prospective Studies ; ROC Curve ; Respiration, Artificial* ; Stroke Volume/physiology*
Keywords
Aortic blood flow velocity ; Children ; Fluid responsiveness ; Noninvasive cardiac output monitor ; Stroke volume variation
Abstract
Continuous noninvasive cardiac output monitoring (NICOM) is a clinically useful tool in the pediatric setting. This study compared the ability of stroke volume variation (SVV) measured by NICOM with that of respiratory variations in the velocity of aortic blood flow (△Vpeak) and central venous pressure (CVP) to predict of fluid responsiveness in mechanically ventilated children after ventricular septal defect repair. The study investigated 26 mechanically ventilated children after the completion of surgery. At 30 min after their arrival in an intensive care unit, a colloid solution of 10 ml/kg was administrated for volume expansion. Hemodynamic variables, including CVP, stroke volume, and △Vpeak in addition to cardiac output and SVV in NICOM were measured before and 10 min after volume expansion. The patients with a stroke volume increase of more than 15 % after volume expansion were defined as responders. The 26 patients in the study consisted of 13 responders and 13 nonresponders. Before volume expansion, △Vpeak and SVV were higher in the responders (both p values <0.001). The areas under the receiver operating characteristic curves of △Vpeak, SVV, and CVP were respectively 0.956 (95 % CI 0.885–1.00), 0.888 (95 % CI 0.764–1.00), and 0.331 (95 % CI 0.123–0.540). This study showed that SVV by NICOM and △Vpeak by echocardiography, but not CVP, reliably predicted fluid responsiveness during mechanical ventilation after ventricular septal defect repair in children.
Full Text
http://link.springer.com/article/10.1007%2Fs00246-013-0772-7
DOI
10.1007/s00246-013-0772-7
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Anesthesiology and Pain Medicine (마취통증의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Ji Young(김지영) ORCID logo https://orcid.org/0000-0001-5822-0338
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/98054