380 633

Cited 0 times in

다양한 교정용 미니 스크류의 인공 피질골 두께에 따른 삽입 토오크와 Pull-out 강도 비교

DC Field Value Language
dc.contributor.author차정열-
dc.contributor.author황충주-
dc.date.accessioned2014-12-21T17:07:53Z-
dc.date.available2014-12-21T17:07:53Z-
dc.date.issued2007-
dc.identifier.issn2234-7518-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/97004-
dc.description.abstractObjective: The purpose of this study was to evaluate the mechanical performance of mini-screws during insertion into artificial bone with use of the driving torque tester (Biomaterials Korea, Seoul, Korea), as well as testing of Pull-out Strength (POS). Methods: Experimental bone blocks with different cortical bone thickness were used as specimens. Three modules of commercially available drill-free type mini-screws (Type A; pure cylindrical type, Biomaterials Korea, Seoul, Korea, Type B; partially cylindrical type, Jeil Medical, Seoul, Korea, Type C; combination type of cylindrical and tapered portions, Ortholution, Seoul, Korea), were used. Results: Difference in the cortical bone thickness had little effect on the maximum insertion torque (MIT) in Type A mini-screws. But in Type B and C, MIT increased as the cortical bone thickness Increased. MIT of Type C was highest in all situations, then Type B and Type A in order. Type C showed lower POS than Type A or B in all situations. There were statistically significant correlations between cortical bone thickness and MIT, and POS for each type of the mini-screws. Conclusion: Since different screw designs showed different insertion torques with increases in cortical bone thickness, the best suitable screw design should be selected according to the different cortical thicknesses at the implant sites.]-
dc.description.statementOfResponsibilityopen-
dc.format.extent5~15-
dc.relation.isPartOfKOREAN JOURNAL OF ORTHODONTICS-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.title다양한 교정용 미니 스크류의 인공 피질골 두께에 따른 삽입 토오크와 Pull-out 강도 비교-
dc.title.alternativeEvaluation of Insertion torque and Pull-out strength of mini-screws according to different thickness of artificial cortical bone-
dc.typeArticle-
dc.contributor.collegeCollege of Dentistry (치과대학)-
dc.contributor.departmentDept. of Orthodontics (교정과학)-
dc.contributor.googleauthor송영윤-
dc.contributor.googleauthor차정열-
dc.contributor.googleauthor황충주-
dc.admin.authorfalse-
dc.admin.mappingfalse-
dc.contributor.localIdA04006-
dc.contributor.localIdA04492-
dc.relation.journalcodeJ02087-
dc.identifier.eissn2005-372X-
dc.contributor.alternativeNameCha, Jung Yul-
dc.contributor.alternativeNameHwang, Chung Ju-
dc.contributor.affiliatedAuthorCha, Jung Yul-
dc.contributor.affiliatedAuthorHwang, Chung Ju-
dc.rights.accessRightsfree-
dc.citation.volume37-
dc.citation.number1-
dc.citation.startPage5-
dc.citation.endPage15-
dc.identifier.bibliographicCitationKOREAN JOURNAL OF ORTHODONTICS, Vol.37(1) : 5-15, 2007-
dc.identifier.rimsid53589-
dc.type.rimsART-
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Orthodontics (교정과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.