VERSANT Hepatitis B Virus DNA 3.0 검사와 Digene Hybrid Capture II Hepatitis B Virus DNA 검사의 비교 및 B형 간염 임상상과의 관련성
Other Titles
Comparison of VERSANT Hepatitis B Virus DNA 3.0 Assay with Digene Hybrid Capture II Hepatitis B Virus DNA Test in Relation to Clinical Status of Hepatitis B Virus
Authors
송은영 ; 허미나 ; 노은연 ; 이경아 ; 최영숙 ; 이창훈 ; 윤여민 ; 이규만
Citation
KOREAN JOURNAL OF LABORATORY MEDICINE , Vol.27(6) : 451-457, 2007
Background :Some differences exist among various Hepatitis B virus (HBV) DNA quantification assays due to lack of standardization and besides clinical usefulness has not been firmly elucidated in Korean HBV patients. Method :We compared Bayer VERSANT HBV DNA 3.0 Assay (VERSANT 3.0) with Digene Hybrid Capture II HBV DNA Test (HC-II) according to HBeAg status and ALT levels in 232 HBV-infected Korean patients. One hundred and seventeen sera with undetectable DNA levels by HC-II were further analyzed by Real-Q HBV quantification assay (BioSewoom). Result :Although VERSANT 3.0 and HC-II showed an excellent correlation (r=0.9739), the results (copies/mL) by VERSANT 3.0 were 0.45 log10 higher than those by HC-II. HBV DNA levels were higher in HBeAg-positive group than in HBeAg-negative group (P=0.002), and in abnormal ALT group than in normal ALT group (P Conclusion :VERSANT 3.0 and HC-II showed an excellent correlation, but a little difference (0.45 log10) existed. VERSANT 3.0 effectively measured clinically relevant HBV DNA levels in most HBVinfected patients in Korea. However, more sensitive assays are needed for patients with negative HBeAg and normal ALT to see the low copies of HBV DNA levels.