2 706

Cited 75 times in

Factors associated with false positives in retinal nerve fiber layer color codes from spectral-domain optical coherence tomography

DC Field Value Language
dc.contributor.author김지현-
dc.contributor.author김찬윤-
dc.contributor.author노승수-
dc.contributor.author성공제-
dc.date.accessioned2014-12-20T16:52:06Z-
dc.date.available2014-12-20T16:52:06Z-
dc.date.issued2011-
dc.identifier.issn0161-6420-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/93530-
dc.description.abstractPURPOSE: To determine the factors that contribute to false-positive retinal nerve fiber layer (RNFL) color code results from spectral-domain optical coherence tomography (OCT). DESIGN: A prospective, cross-sectional study. PARTICIPANTS: This study included 149 eyes from 77 healthy participants. METHODS: Participants, who were consecutively enrolled from June 2009 to December 2009, underwent Cirrus OCT. Recorded demographic and clinical factors included age, gender, eye side, intraocular pressure, central corneal thickness, spherical equivalent, axial length, anterior chamber depth, disc area, and the extent of retinal vasculature. MAIN OUTCOME MEASURES: An abnormal finding in RNFL color codes was defined as ≥1 yellow or red sectors by quadrant and clock-hour maps and a wedge-shaped color pattern represented by yellow or red in the deviation map. The incidence of false-positive color codes was determined. The influence of clinical and demographic factors on the incidence of false-positive RNFL color codes was assessed using generalized linear mixed model analysis. RESULTS: The false-positive rate for ≥1 of the quadrant, clock-hour, and deviation maps was 26.2%. Longer axial length and smaller disc area were significantly associated with an increased incidence of false-positives when other factors were controlled (odds ratios, 2.422 and 0.165; P = 0.008 and 0.035, respectively). CONCLUSIONS: The factors that significantly affected the false-positive RNFL color code results using spectral-domain OCT were axial length and disc area, which may significantly affect the specificity of spectral-domain OCT. Therefore, axial length and disc area should be considered during RNFL thickness profile analysis.-
dc.description.statementOfResponsibilityopen-
dc.format.extent1774~1781-
dc.relation.isPartOfOPHTHALMOLOGY-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.subject.MESHAdult-
dc.subject.MESHCross-Sectional Studies-
dc.subject.MESHFalse Positive Reactions-
dc.subject.MESHFemale-
dc.subject.MESHGlaucoma, Open-Angle/diagnosis*-
dc.subject.MESHGonioscopy-
dc.subject.MESHHumans-
dc.subject.MESHIntraocular Pressure-
dc.subject.MESHMale-
dc.subject.MESHMiddle Aged-
dc.subject.MESHNerve Fibers/pathology*-
dc.subject.MESHPredictive Value of Tests-
dc.subject.MESHProspective Studies-
dc.subject.MESHRetinal Ganglion Cells/pathology*-
dc.subject.MESHRisk Factors-
dc.subject.MESHSensitivity and Specificity-
dc.subject.MESHTomography, Optical Coherence/classification*-
dc.subject.MESHTonometry, Ocular-
dc.titleFactors associated with false positives in retinal nerve fiber layer color codes from spectral-domain optical coherence tomography-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Ophthalmology (안과학)-
dc.contributor.googleauthorNa Rae Kim-
dc.contributor.googleauthorHyunsun Lim-
dc.contributor.googleauthorJi Hyun Kim-
dc.contributor.googleauthorSeung Soo Rho-
dc.contributor.googleauthorGong Je Seong-
dc.contributor.googleauthorChan Yun Kim-
dc.identifier.doi10.1016/j.ophtha.2011.01.058-
dc.admin.authorfalse-
dc.admin.mappingfalse-
dc.contributor.localIdA00998-
dc.contributor.localIdA01035-
dc.contributor.localIdA01284-
dc.contributor.localIdA01946-
dc.relation.journalcodeJ02432-
dc.identifier.eissn1549-4713-
dc.identifier.pmid21550120-
dc.identifier.urlhttp://www.sciencedirect.com/science/article/pii/S016164201100128X-
dc.contributor.alternativeNameKim, Ji Hyun-
dc.contributor.alternativeNameKim, Chan Yun-
dc.contributor.alternativeNameRho, Seung Soo-
dc.contributor.alternativeNameSeong, Gong Je-
dc.contributor.affiliatedAuthorKim, Ji Hyun-
dc.contributor.affiliatedAuthorKim, Chan Yun-
dc.contributor.affiliatedAuthorRho, Seung Soo-
dc.contributor.affiliatedAuthorSeong, Gong Je-
dc.rights.accessRightsnot free-
dc.citation.volume118-
dc.citation.number9-
dc.citation.startPage1774-
dc.citation.endPage1781-
dc.identifier.bibliographicCitationOPHTHALMOLOGY, Vol.118(9) : 1774-1781, 2011-
dc.identifier.rimsid28288-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.