2 473

Cited 40 times in

Comparison of the reliability of "0.5" and "APEX" mark measurements in two frequency-based electronic apex locators

DC Field Value Language
dc.contributor.author이승종-
dc.contributor.author정일영-
dc.date.accessioned2014-12-20T16:33:47Z-
dc.date.available2014-12-20T16:33:47Z-
dc.date.issued2011-
dc.identifier.issn0099-2399-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/92952-
dc.description.abstractINTRODUCTION: A measuring instrument must be both accurate and reliable. This study compared the reliability of the "0.5" and "APEX" mark measurements by using 2 impedance quotient-based electronic apex locators (EALs). METHODS: One hundred four extracted human premolars were used in this study. After access preparation, the teeth were embedded in an alginate model. By using 2 EALs (Root ZX and i-Root), the tooth length was measured at the "0.5" and "APEX" marks with K-files. The file was then cemented, and the apical 3-4 mm was trimmed for the photograph under an operating microscope. The distance between the tip of the file and major foramen (MF) was measured. The intraclass correlation coefficient, the Bland-Altman plot, and box plot were used to compare the reliability. RESULTS: The intraclass correlation coefficient ranged from 0.976-0.994, indicating excellent agreement in both "0.5" and "APEX" marks. The Bland-Altman plots showed that the limits of agreement (mean ± 2 standard deviations) were small enough to confirm that both marks of the 2 EALs can be used for clinical purposes. The distribution of the measurements and outliers was analyzed by using box plots, and it was found that there was no significant difference between the 2 marks. CONCLUSIONS: There was no significant difference in the reliability of the "0.5" and "APEX" marks for locating the MF in both devices. Accordingly, knowing where each mark indicates is more important for determining the working length than which mark to choose.-
dc.description.statementOfResponsibilityopen-
dc.format.extent49~52-
dc.relation.isPartOfJOURNAL OF ENDODONTICS-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.subject.MESHDental Instruments*-
dc.subject.MESHDental Pulp Cavity/anatomy & histology*-
dc.subject.MESHElectric Impedance-
dc.subject.MESHElectronics, Medical/instrumentation-
dc.subject.MESHEquipment Design-
dc.subject.MESHHumans-
dc.subject.MESHOdontometry/instrumentation*-
dc.subject.MESHReproducibility of Results-
dc.subject.MESHRoot Canal Preparation/instrumentation*-
dc.subject.MESHTooth Apex/anatomy & histology*-
dc.titleComparison of the reliability of "0.5" and "APEX" mark measurements in two frequency-based electronic apex locators-
dc.typeArticle-
dc.contributor.collegeCollege of Dentistry (치과대학)-
dc.contributor.departmentDept. of Conservative Dentistry (보존과학)-
dc.contributor.googleauthorIl-Young Jung-
dc.contributor.googleauthorBum-Hee Yoon-
dc.contributor.googleauthorShin-Jae Lee-
dc.contributor.googleauthorSeung Jong Lee-
dc.identifier.doi10.1016/j.joen.2010.08.048-
dc.admin.authorfalse-
dc.admin.mappingfalse-
dc.contributor.localIdA02925-
dc.contributor.localIdA03700-
dc.relation.journalcodeJ01393-
dc.identifier.eissn1878-3554-
dc.identifier.pmid21146076-
dc.identifier.urlhttp://www.sciencedirect.com/science/article/pii/S0099239910007247-
dc.subject.keywordElectronic apex locators-
dc.subject.keywordreliability-
dc.subject.keywordRoot ZX-
dc.subject.keywordworking length-
dc.contributor.alternativeNameLee, Seung Jong-
dc.contributor.alternativeNameJung, Il Young-
dc.contributor.affiliatedAuthorLee, Seung Jong-
dc.contributor.affiliatedAuthorJung, Il Young-
dc.rights.accessRightsnot free-
dc.citation.volume37-
dc.citation.number1-
dc.citation.startPage49-
dc.citation.endPage52-
dc.identifier.bibliographicCitationJOURNAL OF ENDODONTICS, Vol.37(1) : 49-52, 2011-
dc.identifier.rimsid27926-
dc.type.rimsART-
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Conservative Dentistry (보존과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.