Cited 0 times in 
Cited 3 times in 
Stability of commonly used antibiotics solutions in the accufuser® elastometric infusion device under recommended storage and used conditions
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | Lee, Ga-Young | - |
| dc.contributor.author | Kim, Min-Jeong | - |
| dc.contributor.author | Kang, Mina | - |
| dc.contributor.author | Park, Yoo-Sin | - |
| dc.contributor.author | Kim, Shin-Hee | - |
| dc.contributor.author | Kim, Sang-Yeon | - |
| dc.contributor.author | Kang, Ju-Seop | - |
| dc.date.accessioned | 2026-05-14T02:47:47Z | - |
| dc.date.available | 2026-05-14T02:47:47Z | - |
| dc.date.created | 2026-05-13 | - |
| dc.date.issued | 2011-05 | - |
| dc.identifier.issn | 1876-3960 | - |
| dc.identifier.uri | https://ir.ymlib.yonsei.ac.kr/handle/22282913/212190 | - |
| dc.description.abstract | Objectives: The stability of two kinds of solution of penicillin G potassium [PEN-K, 10K U/mL in 5% dextrose water, D5W and 50K U/mL in 0.9% sodium chloride in water, NS) into the intravenous infusion device, Accufuser®, storage were evaluated at controlled temperature (CT, 4±2°C and RT, 25±2°C) during 6 weeks. Methods: Studies were performed using both NS (PEN-K 50K U/mL) and D5W (PEN-K 10K U/mL) injectable solutions. The resulting solutions were transferred to Accufuser® infusion device for storage at CT or RT. Effects of period of storage (from 0 to 6 weeks) and temperature (CT and RT) on the physical appearance and concentrations of PEN-K were determined by visual clarity, pH and antibiotic concentrations by measurement with stability-indicating high-performance liquid chromatography (HPLC) method. Results: In NS (50 K U/mL) and D5W (10K U/mL) solutions, concentrations of PEN-K were slightly change and remained 94.18 and 83.42% at CT after 1 week, respectively. Otherwise, these were rapidly decreased with time and reached to 74.65 and 75.59% in NS (50K U/mL) and D5W (10K U/mL) solutions at RT after 48 h, respectively. Moderate decrement of pH was observed in cold storage and it was shown no significant changes at 6 weeks in the RT conditions. In CT, no significant changes in physical appearance or color of the solutions were observed during the study. Conclusion: Two kinds of PEN-K D5W(10K U/mL) and NS(50K U/mL) solution were shown different chemical stability with time in Accufuser® infusion device without any significant physical change and retained about 70% of initial concentration after 48 h in RT and 80 % after 2 weeks in CT, respectively. We suggested that PEN-K solutions of 50K U/mL NS and 10K U/mL D5W in an Accufuser® infusion device should be preferentially applicable only in CT for 48 h in clinical situations. © Lee et al.; Licensee Bentham Open. | - |
| dc.language | 영어 | - |
| dc.relation.isPartOf | Open Nutraceuticals Journal | - |
| dc.title | Stability of commonly used antibiotics solutions in the accufuser® elastometric infusion device under recommended storage and used conditions | - |
| dc.type | Article | - |
| dc.contributor.googleauthor | Lee, Ga-Young | - |
| dc.contributor.googleauthor | Kim, Min-Jeong | - |
| dc.contributor.googleauthor | Kang, Mina | - |
| dc.contributor.googleauthor | Park, Yoo-Sin | - |
| dc.contributor.googleauthor | Kim, Shin-Hee | - |
| dc.contributor.googleauthor | Kim, Sang-Yeon | - |
| dc.contributor.googleauthor | Kang, Ju-Seop | - |
| dc.identifier.doi | 10.2174/1876396001104010125 | - |
| dc.subject.keyword | Accufuser® | - |
| dc.subject.keyword | Infusion device | - |
| dc.subject.keyword | Penicillin G potassium | - |
| dc.subject.keyword | Stability | - |
| dc.contributor.affiliatedAuthor | Kang, Mina | - |
| dc.identifier.scopusid | 2-s2.0-79958106983 | - |
| dc.citation.volume | 4 | - |
| dc.citation.startPage | 125 | - |
| dc.citation.endPage | 129 | - |
| dc.identifier.bibliographicCitation | Open Nutraceuticals Journal, Vol.4 : 125-129, 2011-05 | - |
| dc.identifier.rimsid | 92889 | - |
| dc.type.rims | ART | - |
| dc.description.journalClass | 1 | - |
| dc.description.journalClass | 1 | - |
| dc.subject.keywordAuthor | Accufuser® | - |
| dc.subject.keywordAuthor | Infusion device | - |
| dc.subject.keywordAuthor | Penicillin G potassium | - |
| dc.subject.keywordAuthor | Stability | - |
| dc.type.docType | Article | - |
| dc.description.isOpenAccess | Y | - |
| dc.description.journalRegisteredClass | scopus | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.