0 0

Cited 0 times in

Cited 0 times in

Hydrodynamic Characteristics of Bovine Pericardial and Porcine Valves Using a Mock Circulatory System Mimicking the Aortic and Pulmonary Positions

Authors
 Shin, Yu Rim  ;  Lee, Seung-Hyun  ;  Shim, Jae-Kwang  ;  Kim, Yongwoo  ;  Lee, Sak 
Citation
 YONSEI MEDICAL JOURNAL, Vol.67(4) : 297-305, 2026-04 
Journal Title
YONSEI MEDICAL JOURNAL
ISSN
 0513-5796 
Issue Date
2026-04
MeSH
Animals ; Aortic Valve* / physiology ; Bioprosthesis* ; Cattle ; Heart Valve Prosthesis* ; Hydrodynamics ; Pericardium* ; Pulmonary Valve* / physiology ; Swine
Keywords
Prosthesis ; heart valve ; pulmonary valve ; bioprostheses
Abstract
Purpose: Aortic prostheses are used in pulmonary positions due to structural similarities between the pulmonary and aortic valves. However, there are no available studies that have comprehensively evaluated the mechanism of bioprosthetic aortic valves under pulmonary conditions. Materials and Methods: Using a mock circulatory system, we evaluated the hydrodynamic characteristics of bovine pericardial and porcine valves. Geometric orifice area, regurgitant and leakage volume, regurgitant fraction, peak pressure gradient, and forward flow volume were evaluated in different pulmonary pressure conditions (from 15/5 mm Hg to 75/35 mm Hg) and normal aortic pressure (110/80 mm Hg). Results: Bovine pericardial valves were associated with larger opening area (0.93 +/- 0.01 vs.1.70 +/- 0.01 for 23-mm valve; 0.99 +/- 0.01 vs.1.75 +/- 0.01 for 25-mm valve; 1.58 +/- 0.01 vs. 2.25 +/- 0.02 for 27-mm valve; all p<0.01) and forward flow volume (42.27 +/- 0.05 vs. 64.79 +/- 0.14 for 23-mm valve; 46.41 +/- 0.06 vs. 64.28 +/- 0.18 for 25-mm valve; 72.64 +/- 0.17 vs.73.25 +/- 0.07 for 27-mm valve; all p<0.01). Porcine valves were associated with incomplete opening, smaller opening area, and lower regurgitant fraction. Bovine pericardial valves demonstrated lower peak pressure gradients (15.75 +/- 0.14 vs. 12.57 +/- 0.47 for 23-mm valve; 14.85 +/- 0.05 vs. 12.87 +/- 0.28 for 25-mm valve; 15.72 +/- 0.32 vs. 7.91 +/- 0.03 for 27-mm valve). Conclusion: Bovine pericardial and porcine bioprosthetic valves has different hydrodynamic characteristics under various pulmonary pressure conditions.
Files in This Item:
92379.pdf Download
DOI
10.3349/ymj.2025.0174
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Thoracic and Cardiovascular Surgery (흉부외과학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Anesthesiology and Pain Medicine (마취통증의학교실) > 1. Journal Papers
Yonsei Authors
Shin, Yu Rim(신유림)
Shim, Jae Kwang(심재광) ORCID logo https://orcid.org/0000-0001-9093-9692
Lee, Sak(이삭) ORCID logo https://orcid.org/0000-0001-6130-2342
Lee, Seung Hyun(이승현)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/211827
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links