Comparing the Biomechanical Stability of Cerclage Cable with Plate Insert Versus Locking Screw in Periprosthetic Humeral Fracture

Authors
 Yong-Jun Lee  ;  Joon-Ryul Lim  ;  Jee-Hoon Choi  ;  Tae-Hwan Yoon  ;  Yun-Rak Choi  ;  Yong-Min Chun 
Citation
 JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, Vol.106(9) : 817-822, 2024-05 
Journal Title
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME
ISSN
 0021-9355 
Issue Date
2024-05
MeSH
Aged ; Aged, 80 and over ; Biomechanical Phenomena ; Bone Plates* ; Bone Screws* ; Cadaver* ; Female ; Fracture Fixation, Internal* / adverse effects ; Fracture Fixation, Internal* / instrumentation ; Fracture Fixation, Internal* / methods ; Humans ; Humeral Fractures / surgery ; Male ; Periprosthetic Fractures* / etiology ; Periprosthetic Fractures* / surgery
Abstract
BACKGROUND: In the setting of periprosthetic humeral fractures, the humeral stem of the implant represents a substantial challenge to the optimal method of proximal fixation. This study aimed to compare the initial biomechanical stability provided by cerclage cables with a locking plate insert versus bicortical locking screws (i.e., the gold standard for fixation) in fresh cadaveric humeri. METHODS: After calculating the sample size, we utilized 10 sets of cadaveric specimens and created a 5-mm osteotomy gap 120 mm distal to the tip of the greater tuberosity, simulating a Wright and Cofield type-B periprosthetic humeral fracture on each specimen. Using 3 locking screws for distal fragment fixation, identical in all specimens, the specimens were assigned to Group A (3 cerclage cables with a plate insert) or Group B (3 locking bicortical screws) for proximal fragment fixation. Biomechanical tests included stiffness in varus and valgus bending, torsion, and axial compression, and a single load to failure. RESULTS: No significant differences were observed in the biomechanical metrics between the 2 groups. CONCLUSIONS: Our study revealed that fixation with use of cerclage cables with a plate insert demonstrated biomechanical stability comparable with that of bicortical locking screw fixation when addressing the proximal fragment in Wright and Cofield type-B periprosthetic humeral fractures. CLINICAL RELEVANCE: For proximal fragment fixation of periprosthetic humeral fractures, cerclage cables with a plate insert can be utilized as an effective fixation method that offers initial fixation strength that is comparable to the use of 3 locking bicortical screws.
Full Text
https://journals.lww.com/jbjsjournal/fulltext/2024/05010/comparing_the_biomechanical_stability_of_cerclage.9.aspx
DOI
10.2106/JBJS.23.00830
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Orthopedic Surgery (정형외과학교실) > 1. Journal Papers
Yonsei Authors
Yoon, Tae Hwan(윤태환) ORCID logo https://orcid.org/0000-0002-2859-5240
Lee, Yong Joon(이용준)
Lim, Joon Ryul(임준열) ORCID logo https://orcid.org/0000-0002-0123-7136
Chun, Yong Min(천용민) ORCID logo https://orcid.org/0000-0002-8147-6136
Choi, Yun Rak(최윤락)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/201024
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links