Axial length measurements: Comparison of a new swept-source optical coherence tomography-based biometer and partial coherence interferometry in myopia

Authors
 Jong Yun Yang  ;  Hong Kyu Kim  ;  Sung Soo Kim 
Citation
 JOURNAL OF CATARACT AND REFRACTIVE SURGERY, Vol.43(3) : 328-332, 2017-03 
Journal Title
JOURNAL OF CATARACT AND REFRACTIVE SURGERY
ISSN
 0886-3350 
Issue Date
2017-03
MeSH
Anterior Chamber ; Axial Length, Eye ; Biometry ; Humans ; Interferometry* ; Myopia* / diagnostic imaging ; Prospective Studies ; Tomography, Optical Coherence*
Abstract
Purpose: To compare axial length (AL) measurements between a swept-source optical coherence tomography (SS-OCT) biometer (IOLMaster 700) and a partial coherence interferometry (PCI) biometer (IOLMaster, version 5.4) in myopic eyes.

Setting: Severance Hospital, Yonsei University, Seoul, South Korea.

Design: Prospective evaluation of a diagnostic test.

Methods: This study enrolled patients with a diagnosis of myopia in their medical records. Two consecutive AL measurements were performed in each eye in random order using the 2 biometers. Subanalysis was performed according to lens status, fixation status, degree of myopia, and the presence of posterior staphyloma. The interdevice agreement was evaluated with Bland-Altman analyses and paired t tests.

Results: This study comprised 219 eyes of 117 patients. During a fixation check using SS-OCT, 19.6% of eyes showed fixation loss. Overall, the SS-OCT biometer showed a longer AL than the PCI biometer. (P < .001) In a subanalysis that categorized eyes according to lens status, fixation status, degree of myopia, and the presence of posterior staphyloma, the SS-OCT biometer also showed longer ALs than the PCI biometer. Of eyes with good fixation, significant differences in AL measurements between devices were detected for those with posterior staphyloma (P < .001) but not for those without posterior staphyloma (P = .104).

Conclusions: The 2 devices showed differences in AL measurements in myopic eyes. Fixation status and the presence of posterior staphyloma were important factors in these differences. For myopic eyes with posterior staphyloma, the SS-OCT biometer is expected to produce more precise AL measurements because it allows evaluation of the fixation status.
Full Text
https://journals.lww.com/jcrs/Fulltext/2017/03000/Axial_length_measurements__Comparison_of_a_new.7.aspx
DOI
10.1016/j.jcrs.2016.12.023
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Sung Soo(김성수) ORCID logo https://orcid.org/0000-0002-0574-7993
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/195748
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links