63 135

Cited 0 times in

In vitro pressure drop comparison between two mechanical valve prostheses

Authors
 Sang Hyun Kim  ;  Byung Chul Chang  ;  Bum Koo Cho 
Citation
 YONSEI MEDICAL JOURNAL, Vol.35(1) : 72-76, 1994-03 
Journal Title
YONSEI MEDICAL JOURNAL
ISSN
 0513-5796 
Issue Date
1994-03
MeSH
Heart Valve Prosthesis* ; Models, Cardiovascular ; Models, Structural ; Pressure
Abstract
An hemodynamic evaluation of two mechanical heart valves is presented. A tilting disc valve and a bileaflet valve were incorporated in a mock circulatory system which consists of a closed flow loop with a pneumatically driven flexible diaphragm to simulate the physiologic pulsatile flow. Comparisons between the valves were made on the aortic pressure, ventricular pressure, as well as mean pressure gradient at a systolic duration of 45% and a heart rate of 90 beats per minute. The results showed that the tilting disc valve has higher ventricular pressure and mean pressure gradient than that of the bileaflet valve. This indicates that the tilting disc valve has higher transvalvular flow resistance and energy loss than that of the bileaflet valve. From this study it is demonstrated that the mock circulatory system can be a very useful device to evaluate the prosthetic heart valves in vitro.
Files in This Item:
T199400685.pdf Download
DOI
10.3349/ymj.1994.35.1.72
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Thoracic and Cardiovascular Surgery (흉부외과학교실) > 1. Journal Papers
Yonsei Authors
Chang, Byung Chul(장병철)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/194950
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links