53 135

Cited 0 times in

In vitro sound spectral analysis of prosthetic heart valves by mock circulatory system

Authors
 Sang Hyun Kim  ;  Byung Chul Chang  ;  Gyerae Tack  ;  Jae Man Huh  ;  Meyun Shick Kang  ;  Bum Koo Cho  ;  Young Hwan Park 
Citation
 YONSEI MEDICAL JOURNAL, Vol.35(3) : 271-278, 1994-09 
Journal Title
YONSEI MEDICAL JOURNAL
ISSN
 0513-5796 
Issue Date
1994-09
MeSH
Fourier Analysis ; Heart Sounds* ; Heart Valve Prosthesis* ; Humans ; Models, Cardiovascular* ; Signal Processing, Computer-Assisted
Abstract
A comparative study was made of the sounds produced by normal prosthetic valves (St. Jude Medical, Bjork-Shiley, polymer) with those produced by the same valves but having simulated thrombosis at the stent, hinge, or strut. Comparisons of the closing sound were made for the power frequency spectra associated with individual valves. We used periodogram approach to obtain the spectral characteristics of the valve prostheses. The closing sound of the abnormal mechanical valves displayed lower apparent peak frequency. But the abnormal polymer valve produced higher apparent peak frequency. The results showed that frequency spectra gave information pertinent to the simulated malfunction. Sound spectral analysis is believed to be a simple and a good diagnostic tool for detection of prosthetic valve malfunction. Also it seemed to be superior to other methods such as phonocardiography and echocardiography.
Files in This Item:
T199400523.pdf Download
DOI
10.3349/ymj.1994.35.3.271
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Thoracic and Cardiovascular Surgery (흉부외과학교실) > 1. Journal Papers
Yonsei Authors
Chang, Byung Chul(장병철)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/194898
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links