A Novel Computed Tomography Image Reconstruction for Improving Visualization of Pulmonary Vasculature: Comparison Between Preprocessing and Postprocessing Images Using a Contrast Enhancement Boost Technique

Authors
 Chuluunbaatar Otgonbaatar  ;  Jae-Kyun Ryu  ;  Hackjoon Shim  ;  Pil-Hyun Jeon  ;  Sang-Hyun Jeon  ;  Jin Woo Kim  ;  Sung Min Ko  ;  Hyunjung Kim 
Citation
 JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, Vol.46(5) : 729-734, 2022-09 
Journal Title
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY
ISSN
 0363-8715 
Issue Date
2022-09
MeSH
Angiography ; Contrast Media* ; Humans ; Image Processing, Computer-Assisted / methods ; Signal-To-Noise Ratio ; Tomography, X-Ray Computed* / methods
Abstract
Objective: This study aimed to evaluate chest computed tomography (CT) angiography image quality using the contrast enhancement (CE)-boost technique compared with conventional images.

Methods: Forty patients who underwent contrast-enhanced chest CT were included. Combined CT angiography images of the iodinated image obtained from the subtraction of nonenhanced CT images and CT angiography images were used to generate CE-boost images. Computed tomography attenuation, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) for the right and left pulmonary arteries as the central and subsegmental arteries as peripheral vessels were assessed. Subjective image quality was rated on a 5-point scale by 2 radiologists. Image quality was assessed using a paired t test.

Results: Computed tomography attenuation in the main pulmonary artery was significantly higher for the CE-boost images (311.05 ± 91.94) than for the conventional images (221.25 ± 61.21, P < 0.001). Similarly, the CE-boost images resulted in significantly higher CT attenuation in the subsegmental arteries (right, 305.34 ± 90.13; left, 313.05 ± 97.21) than in the conventional images (right, 218.45 ± 63.16; left, 223.89 ± 74.27). The CE-boost technique demonstrated marked improvement in the visualization of the peripheral pulmonary artery without the administration of a higher iodine delivery rate. The mean SNR and CNR were also significantly higher in the central and peripheral vessels in the CE-boost images than in the conventional images (P < 0.001). In the subjective analysis, the image contrast and vascular contrast edge were significantly higher for the CE-boost images than for conventional images (P < 0.001).

Conclusions: The CE-boost technique increases not only the visualization of peripheral arteries by improving vascular attenuation but also the SNR and CNR.
Full Text
https://journals.lww.com/jcat/Fulltext/2022/09000/A_Novel_Computed_Tomography_Image_Reconstruction.10.aspx
DOI
10.1097/RCT.0000000000001347
Appears in Collections:
1. College of Medicine (의과대학) > Yonsei Biomedical Research Center (연세의생명연구원) > 1. Journal Papers
Yonsei Authors
Shim, Hack Joon(심학준)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/193404
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links