Background: Pediatric cancer patients undergoing chemotherapy or radiation therapy generally require a central venous catheter (CVC). However, serum drawn from CVCs has several drawbacks for use in routine chemistry tests. Biochemical analytes were evaluated using heparin plasma instead of serum to maintain turnaround time and to prevent problems caused by micro-clot formation or delayed clotting time.

Methods: Venous blood samples from 52 pediatric oncology patients with chemoports or Hickman catheters were collected in serum separating tubes (SSTs) and lithium heparin tubes (LHTs). A total of 29 parameters were analyzed on a Cobas c702 (Roche Diagnostics, Mannheim, Germany). Passing-Bablok regression and Bland-Altman difference plots were used for statistical analyses.

Results: When the mean value of each analyte measured from LHT was compared with those from SST, percentage bias was within the desirable bias limit in most of the analytes. However, albumin, potassium, and inorganic phosphorus showed a negative constant bias of -3.0%, -5.3%, and -1.6%, respectively, and total protein showed a positive constant bias of + 3.8%.

Conclusions: The use of LHTs for sample collection from pediatric patients with CVCs could be helpful for routine chemistry analyses. The results of potassium and total protein should be interpreted with consideration of the difference between serum and plasma samples.
Full Text
https://www.sciencedirect.com/science/article/pii/S0009912022001291
DOI
10.1016/j.clinbiochem.2022.04.017
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > Dept. of Pediatrics (소아과학교실) > 1. Journal Papers
Yonsei Authors
Kang, Hyein(강혜인)
Lyu, Chuhl Joo(유철주) ORCID logo https://orcid.org/0000-0001-7124-7818
Lee, Sang-Guk(이상국) ORCID logo https://orcid.org/0000-0003-3862-3660
Rim, John Hoon(임정훈) ORCID logo https://orcid.org/0000-0001-6825-8479
Lim, Jong Baeck(임종백) ORCID logo https://orcid.org/0000-0003-0419-0422
Hahn, Seung Min(한승민) ORCID logo https://orcid.org/0000-0001-9832-6380
Han, Jung Woo(한정우) ORCID logo https://orcid.org/0000-0001-8936-1205
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/192810
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links