0 173

Cited 0 times in

Performance assessment of ASTA MicroIDSys, a new matrix assisted laser desorption ionization-time of flight mass spectrometry system, for identification of viridans group streptococci

Authors
 O-Jin Lee  ;  Young Jin Ko  ;  Seul-Bi Lee  ;  Choon Mee Kim  ;  Sook-Jin Jang  ;  Joong-Ki Kook  ;  Yun Kyong Lim  ;  Jong Hee Shin  ;  Myung Geun Shin  ;  Seung Jung Kee  ;  Seok Hoon Jeong  ;  Seong-Ho Kang  ;  Geon Park 
Citation
 MICROBIOLOGY AND IMMUNOLOGY, Vol.65(12) : 566-574, 2021-12 
Journal Title
MICROBIOLOGY AND IMMUNOLOGY
ISSN
 0385-5600 
Issue Date
2021-12
MeSH
Bacteria* ; Lasers ; Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization ; Viridans Streptococci*
Keywords
comparative study ; database ; matrix-assisted laser desorption-ionization ; viridans streptococci
Abstract
The performance of the ASTA MicroIDSys system (ASTA), a new matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) system, was evaluated for the identification of viridans group streptococci (VGS) and compared with the results obtained with the Bruker Biotyper system (Bruker Daltonics). A total of 106 Streptococcus reference strains belonging to 24 species from the bacterial strain bank was analyzed using the two MALDI-TOF MS systems. Of the 106 reference strains tested, ASTA MicroIDSys and Bruker Biotyper correctly identified 84.9% and 81.1% at the species level, 100% and 97.2% at the group level and 100% and 98.1% at the genus level, respectively. The difference between the two systems was not statistically significant (P = 0.289). Out of 24 species, 13 species were accurately identified to the species level with 100% accurate identification rates with both systems. The accurate identification rates at the species level of ASTA MicroIDSys and Bruker Biotyper were 100% and 87.5% for the S. anginosus group; 78.4% and 73.5% for the S. mitis group; 91.7% and 91.7% for the S. mutans group; and 100% and 100% for the S. salivarius group, respectively. The ASTA MicroIDSys showed an identification performance equivalent to that of the Bruker Biotyper for VGS. Therefore, it would be useful for the identification of VGS strains in clinical microbiology laboratories.
Full Text
https://onlinelibrary.wiley.com/doi/10.1111/1348-0421.12942
DOI
10.1111/1348-0421.12942
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Jeong, Seok Hoon(정석훈) ORCID logo https://orcid.org/0000-0001-9290-897X
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/191108