A preclinical trial of perventricular pulmonary valve implantation: Pericardial versus aortic porcine valves mounted on self-expandable stent

Authors
 Min-Seok Kim  ;  Whal Lee  ;  Ki-Bum Kim  ;  Hong-Gook Lim  ;  Yong Jin Kim 
Citation
 ARTIFICIAL ORGANS, Vol.45(5) : E89-E100, 2021-05 
Journal Title
ARTIFICIAL ORGANS
ISSN
 0160-564X 
Issue Date
2021-05
MeSH
Animals ; Bioprosthesis* ; Cardiac Catheterization / instrumentation* ; Heart Valve Prosthesis Implantation / instrumentation* ; Heart Valve Prosthesis Implantation / methods ; Heart Valve Prosthesis* ; Hemodynamics ; Models, Animal ; Prosthesis Design ; Pulmonary Valve* ; Self Expandable Metallic Stents ; Sheep ; Swine
Keywords
animal study ; congenital heart disease ; pulmonary valve replacement ; tissue engineering
Abstract
Perventricular pulmonary valve implantation (PPVI) of a xenograft valve can be a less invasive technique that avoids cardiopulmonary bypass in patients who require pulmonary valve replacement. We compared the hemodynamics, durability, and histologic changes between two different xenogenic valves (pericardial vs. aortic valve porcine xenografts) implanted into the pulmonary valve position using a PPVI technique and evaluated the safety and efficacy of PPVI as a preclinical study. In 18 sheep, pericardial (group porcine pericardial [PP], n = 9) or aortic valve (group porcine aortic valve [PAV], n = 9) xenogenic porcine valves manufactured as a stented valve were implanted using a PPVI technique. The porcine tissues were decellularized, alpha-galactosidase treated, fixed with glutaraldehyde after space-filler treatment, and detoxified to improve durability. Hemodynamic and immunohistochemical studies were performed after the implantation; radiologic and histologic studies were performed after a terminal procedure. All stented valves were positioned properly after the implantation, and echocardiography and cardiac catheterization demonstrated good hemodynamic state and function of the valves. All the anti-α-Gal IgM and IgG titers were below 0.3 optical density. Computed tomography of extracted valves demonstrated no significant differences in the degree of calcification between the two groups (P = .927). Microscopic findings revealed a minimal amount of calcification and no significant infiltration of macrophage or T-cell in both groups, regardless of the implantation duration. The PPVI is a feasible technique. Both stented valves made of PP and PAV showed no significant differences in hemodynamic profile, midterm durability, and degree of degenerative dystrophic calcification.
Full Text
https://onlinelibrary.wiley.com/doi/10.1111/aor.13845
DOI
10.1111/aor.13845
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Thoracic and Cardiovascular Surgery (흉부외과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Min-Seok(김민석)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/190918
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links