280 455

Cited 0 times in

Accuracy and Feasibility of a Zero-Setup Implant Guide System Made of a Light-Cured Composite Resin with Simultaneous Flapless Sinus Augmentation: A Pilot Study

DC Field Value Language
dc.contributor.author박진영-
dc.contributor.author백정원-
dc.contributor.author이중석-
dc.contributor.author정의원-
dc.contributor.author차재국-
dc.contributor.author최성호-
dc.date.accessioned2021-12-28T17:51:51Z-
dc.date.available2021-12-28T17:51:51Z-
dc.date.issued2021-08-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/187306-
dc.description.abstractThe aim of this study was to evaluate the accuracy and feasibility of flapless implant placement using a light-cured surgical template made of composite resin comprising dimethacrylate and diurethane, performed simultaneously with crestal sinus augmentation. This pilot study was a prospective, single-armed, controlled clinical trial. Four patients having an edentulous region in the posterior maxilla with 4 to 8 mm of residual ridge height were enrolled. Flapless fully-guided implant placement was performed using a zero-setup surgical template with simultaneous crestal sinus graft. Horizontal, vertical, and angular deviations of the placed implants from the planned positions were measured. Clinical observations were made, and implant stability was measured up to 3 months postoperatively. All four implants were placed and maintained successfully without complication. The mean deviations of the four implants were 1.27 and 1.85 mm horizontally at platform and apex, respectively, 0.84 mm vertically at apex, and 4.76° in angle. Satisfactory implant stability was obtained in all implants at placement and 3 months. Fully-guided implant placement and crestal sinus graft using a zero-setup surgical template can be a safe and feasible method with acceptable accuracy. A randomized controlled trial should be performed in the future to validate its clinical performance in terms of surgical time, accuracy, and patient-reported outcomes.-
dc.description.statementOfResponsibilityopen-
dc.languageEnglish-
dc.publisherMDPI AG-
dc.relation.isPartOfAPPLIED SCIENCES-BASEL-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.titleAccuracy and Feasibility of a Zero-Setup Implant Guide System Made of a Light-Cured Composite Resin with Simultaneous Flapless Sinus Augmentation: A Pilot Study-
dc.typeArticle-
dc.contributor.collegeCollege of Dentistry (치과대학)-
dc.contributor.departmentDept. of Periodontics (치주과학교실)-
dc.contributor.googleauthorJin-Young Park-
dc.contributor.googleauthorJoo-Yeon Lee-
dc.contributor.googleauthorJoo-Nyeon Kim-
dc.contributor.googleauthorJeong-Won Paik-
dc.contributor.googleauthorJung-Seok Lee-
dc.contributor.googleauthorUi-Won Jung-
dc.contributor.googleauthorSeong-Ho Choi-
dc.contributor.googleauthorJae-Kook Cha-
dc.identifier.doi10.3390/app11178085-
dc.contributor.localIdA04749-
dc.contributor.localIdA01836-
dc.contributor.localIdA03185-
dc.contributor.localIdA03692-
dc.contributor.localIdA04004-
dc.contributor.localIdA04081-
dc.relation.journalcodeJ03706-
dc.identifier.eissn2076-3417-
dc.subject.keywordguided implant surgery-
dc.subject.keywordsinus augmentation-
dc.subject.keywordflapless surgery-
dc.contributor.alternativeNamePark, Jin Young-
dc.contributor.affiliatedAuthor박진영-
dc.contributor.affiliatedAuthor백정원-
dc.contributor.affiliatedAuthor이중석-
dc.contributor.affiliatedAuthor정의원-
dc.contributor.affiliatedAuthor차재국-
dc.contributor.affiliatedAuthor최성호-
dc.citation.volume11-
dc.citation.number17-
dc.citation.startPage8085-
dc.identifier.bibliographicCitationAPPLIED SCIENCES-BASEL, Vol.11(17) : 8085, 2021-08-
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Periodontics (치주과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.