103 219

Cited 0 times in

다엽 콜리메이터(Multileaf Collimator)와 합금납 차폐물(Lead Alloy Blok)의 반 그림자영역의 선량 분포상의 특성 비교

Other Titles
 Dosimetric Characteristics on Penumbra Regions of the Multileaf Collimator as Compared with the Lead Alloy Block 
Authors
 이상욱  ;  오영택  ;  김우철  ;  금기창  ;  윤성익  ;  김현수  ;  박원  ;  추성실  ;  김귀언 
Citation
 Journal of the Korean Society of Therapeutics Radiology (대한치료방사선과학회지), Vol.13(4) : 391-396, 1995-12 
Journal Title
Journal of the Korean Society of Therapeutics Radiology(대한치료방사선과학회지)
ISSN
 1225-6765 
Issue Date
1995-12
Abstract
Purpose: The Conformal Radiation Therapy has been widely used under favour of development of computer technologies. The delivery of a large number of static radiation fields are being necessary for the conformal irradiation. In this paper, we investigate dosimetric characteristics on penumbra regions of a multileaf collimator(MLC), and compare to those of lead alloy block for he optimal use of the system in 3-D conformal radiotherapy.

Materials and Methods: The measurement of penumbra by MLC or lead alloy block was performed with 6 or 10 MV X-rays. The film was positioned at a dmax depth and 10 cm depth, and its optical density was determined using a scanning videodensitometer. The effective penumbra, the distance from 80% to 20% isodose lines and 90 to 10 were analyzed as a function of the angle between the direction of leaf motion and the edge defined by leaves.

Results: Increasing MLC angle (0-75°) was observed with increasing the penumbra widths and the scalloping effect. There was no definite differences of penumbra width from 80% to 20% isodose lines, while being the small increase of penumbra width of lead alloy block are agree reasonably with those of MLC within 4.8mm.

Conclusion: The comparative qualitative study of the penumbra between MLC and lead alloy block demonstrate the clinical acceptability and suitability of the multileaf collimator for 3-D conformal radiotherapy.
Files in This Item:
T199500975.pdf Download
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers
Yonsei Authors
Keum, Ki Chang(금기창) ORCID logo https://orcid.org/0000-0003-4123-7998
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/186234
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links