Influence of shoulder coverage difference of abutment on stress distribution and screw stability in tissue-level internal connection implants: A finite element analysis and in vitro study

Authors
 Hag-Young Lee  ;  Seung-Won Yang  ;  You-Jung Kang  ;  Kyung Chul Oh  ;  Jee-Hwan Kim 
Citation
 JOURNAL OF PROSTHETIC DENTISTRY, Vol.125(4) : 682.e1-682.e10, 2021-04 
Journal Title
JOURNAL OF PROSTHETIC DENTISTRY
ISSN
 0022-3913 
Issue Date
2021-04
MeSH
Bone Screws ; Dental Abutments ; Dental Implant-Abutment Design* ; Dental Implants* ; Dental Stress Analysis ; Finite Element Analysis ; Torque
Abstract
Statement of problem: Tissue-level internal connection implants are widely used, but the difference in abutment screw stability because of the shoulder coverage formed by the contact between the shoulder of the implant collar and the abutment remains unclear.

Purpose: The purpose of this finite element analysis (FEA) and in vitro study was to investigate stress distribution and abutment screw stability as per the difference in shoulder coverage of the abutment in tissue-level internal connection implants.

Material and methods: Abutments were designed in 3 groups as per the shoulder coverage of the implant collar, yielding complete coverage (complete group), half coverage (half group), no coverage (no group) groups. In the FEA, a tightening torque of 30.0 Ncm was applied to the abutment screw, a force of 250 N was applied to the crown at a 30-degree angle, and the von Mises stresses and the stress distribution patterns were evaluated. In the in vitro study, the groups were tested (n=12). A total of 200 000 cyclic loads were applied at 250 N, 14 Hz, and at a 30-degree angle. Removal torque values and scanning electron microscopy (SEM) images were assessed. Removal torque values were analyzed by ANOVA and paired t tests.

Results: The maximum von Mises stress of the abutment screw was the lowest in the complete group, slightly higher in the half group, and highest in the no group. High stresses were concentrated in 1 location in the implant abutment connection area of the no group. The removal torque values after loading were significantly lower in the no group than in the complete group (P=.047). The SEM images revealed concentrated structural loss and wear in 1 location of the no group.

Conclusions: FEA and in vitro studies confirmed that the shoulder coverage of the abutment in the tissue-level internal connection implant helped improve screw stability. Cyclic loading reduced the removal torque of the abutment screw.
Full Text
https://www.sciencedirect.com/science/article/pii/S0022391320305618
DOI
10.1016/j.prosdent.2020.08.022
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Prosthodontics (보철과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Jee Hwan(김지환) ORCID logo https://orcid.org/0000-0002-0872-4906
Oh, Kyung Chul(오경철) ORCID logo https://orcid.org/0000-0003-4584-2597
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/184313
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links