332 359

Cited 27 times in

Trueness of ten intraoral scanners in determining the positions of simulated implant scan bodies

Authors
 Ryan Jin Young Kim  ;  Goran I Benic  ;  Ji-Man Park 
Citation
 SCIENTIFIC REPORTS, Vol.11(1) : 2606, 2021-01 
Journal Title
SCIENTIFIC REPORTS
Issue Date
2021-01
Abstract
Few investigations have evaluated the 3-dimensional (3D) accuracy of digital implant scans. The aim of this study was to evaluate the performance of 10 intraoral scanners (IOSs) (CEREC Omnicam, CEREC Primescan, CS 3600, DWIO, i500, iTero Element, PlanScan, Trios 2, Trios 3, and True Definition) in obtaining the accurate positions of 6 cylinders simulating implant scan bodies. Digital scans of each IOS were compared with the reference dataset obtained by means of a coordinate measuring machine. Deviation from the actual positions of the 6 cylinders along the XYZ axes and the overall 3D deviation of the digital scan were calculated. The type of IOSs and position of simulated cylindrical scan bodies affected the magnitude and direction of deviations on trueness. The lowest amount of deviation was found at the cylinder next to the reference origin, while the highest deviation was evident at the contralateral side for all IOSs (p < 0.001). Among the tested IOSs, the CEREC Primescan and Trios 3 had the highest trueness followed by i500, Trios 2, and iTero Element, albeit not statistically significant (p > 0.05), and the DWIO and PlasScan had the lowest trueness in partially edentulous mandible digital implant scans (p < 0.001).
Files in This Item:
T202100576.pdf Download
DOI
10.1038/s41598-021-82218-z
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Prosthodontics (보철과학교실) > 1. Journal Papers
Yonsei Authors
Park, Ji-Man(박지만) ORCID logo https://orcid.org/0000-0003-0018-1166
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/182193
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.