Feasibility and accuracy of a novel automated three-dimensional ultrasonographic analysis system for abdominal aortic aneurysm: comparison with two-dimensional ultrasonography and computed tomography

Authors
 In-Jeong Cho  ;  Jinyong Lee  ;  Jinki Park  ;  Sang-Eun Lee  ;  Chul-Min Ahn  ;  Young-Guk Ko  ;  Donghoon Choi  ;  Hyuk-Jae Chang 
Citation
 CARDIOVASCULAR ULTRASOUND, Vol.18(1) : 24, 2020-07 
Journal Title
CARDIOVASCULAR ULTRASOUND
Issue Date
2020-07
Keywords
Abdominal aortic aneurysm ; Software validation ; Three-dimensional imaging
Abstract
Background: Accurate measurement of the maximum aortic diameter (Dmax) is crucial for patients with abdominal aortic aneurysm (AAA). Aortic computed tomography (CT) provides accurate Dmax values by three-dimensional (3-D) reconstruction but may cause nephrotoxicity because of contrast use and radiation hazard. We aimed to evaluate the accuracy of a novel semi-automated 3-D ultrasonography (3-D US) system compared with that of CT as a reference.

Methods: Patients with AAA (n = 59) or individuals with normal aorta (n = 18) were prospectively recruited in an outpatient setting. Two-dimensional ultrasonography (2-D US) and 3-D US images were acquired with a single-sweep volumetric transducer. The analysis was performed offline with a software. Dmax and the vessel area of the Dmax slice were measured with 2-D US, 3-D US, and CT. The lumen and thrombus areas of the Dmax slice were also measured in 40 patients with intraluminal thrombus. Vessel and thrombus volumes were measured using 3-D US and CT.

Results: The Dmax values from 3-D US demonstrated better agreement (R2 = 0.984) with the CT values than with the 2-D US values (R2 = 0.938). Overall, 2-D US underestimated Dmax compared with 3-D US (32.3 ± 12.1 mm vs. 35.1 ± 12.0 mm). The Bland-Altman analysis of the 3-D US values, revealed better agreement with the CT values (2 standard deviations [SD], 2.9 mm) than with the 2-D US values (2 SD, 5.4 mm). The vessel, lumen, and thrombus areas all demonstrated better agreement with CT than with 2-D US (R2 = 0.986 vs. 0.960 for the vessel, R2 = 0.891 vs. 0.837 for the lumen, and R2 = 0.977 vs. 0.872 for the thrombus). The thrombus volume assessed with 3-D US showed good correlation with the CT value (R2 = 0.981 and 2 SD in the Bland-Altman analysis: 13.6 cm3).

Conclusions: Our novel semi-automated 3-D US analysis system provides more accurate Dmax values than 2-D US and provides precise volumetric data, which were not evaluable with 2-D US. The application of the semi-automated 3-D US analysis system in abdominal aorta assessment is easy and accurate.
Files in This Item:
T202002842.pdf Download
DOI
10.1186/s12947-020-00207-0
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers
Yonsei Authors
Ko, Young Guk(고영국) ORCID logo https://orcid.org/0000-0001-7748-5788
Ahn, Chul-Min(안철민) ORCID logo https://orcid.org/0000-0002-7071-4370
Chang, Hyuk-Jae(장혁재) ORCID logo https://orcid.org/0000-0002-6139-7545
Choi, Dong Hoon(최동훈) ORCID logo https://orcid.org/0000-0002-2009-9760
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/179492
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links