105 240

Cited 0 times in

국산 제2세대 혈소판보존용 혈액백의 평가

Other Titles
 Evaluation of a domestic second generation platelet storage container. 
Authors
 한규섭  ;  김문정  ;  김현옥 
Citation
 Korean Journal of Blood Transfusion (대한수혈학회지), Vol.9(2) : 253-258, 1998 
Journal Title
Korean Journal of Blood Transfusion(대한수혈학회지)
ISSN
 1226-9336 
Issue Date
1998
Abstract
Background : Polyvinyl(PVC) plastic container plasticized with di-(2-ethylhexyl) phthalate(DEHP) has been used for the storage of platelet concentrates for five days in Korea. Authors evaluated a second generation platelet storage container plasticized with tri(2-ethylhexyl) phthalate(TOTM) which was recently produced by Green Cross Medical Corp.(Korea). Methods : 30 units of platelet concentrates were stored in TOTM-PVC container at 22' C in a flatbed agitator Samples were taken at day 1,3,5, and 7 from the containers and tested for platelet count, MPV, PDW, pH, HCO3-,LDH, lactate, hypotonic shock response and β-thromboglobulin(β-TG). Electron microscopic examination was also performed. Results : The number and functions of platelets were well preserved during storage. pH was maintained above 6.8 and any evidence for platelet activation was minimal. Conclusion : The TOTM-PVC second generation platelet storage container recently produced by the Green Cross Medical Corp.(Korea) was able to preserve platelets for at least five days without significant storage lesions.
Files in This Item:
T199802828.pdf Download
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Hyun Ok(김현옥) ORCID logo https://orcid.org/0000-0002-4964-1963
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/176953
사서에게 알리기
  feedback