0 525

Cited 8 times in

Analytical and clinical performance of newly developed immunoassay for detecting thyroid-stimulating immunoglobulin, the Immulite TSI assay

Authors
 Jin Ju Kim  ;  Soon-Ho Jeong  ;  Byungkwang Kim  ;  Dokyun Kim  ;  Seok Hoon Jeong 
Citation
 SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, Vol.79(6) : 443-448, 2019 
Journal Title
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION
ISSN
 0036-5513 
Issue Date
2019
Keywords
Graves’ disease ; Thyrotropin ; hyperthyroidism ; immunoassay ; thyroid-stimulating antibody
Abstract
Graves' disease (GD) is caused by autoantibodies against the thyrotropin receptor (TRAb), and among the three types of TRAbs, only the stimulating type (TSI) is known to be associated with GD. In this study, we evaluated the analytical performance of a new fully automated chemiluminescent TSI immunoassay, namely, the Immulite TSI assay, and compared the diagnostic efficacy of the assay with the Elecsys Anti-TSH receptor (TSHR) assay. Precision was evaluated using two levels of quality control reagents, and linearity was evaluated across the expected analytical measurement range (0.18-37.35 IU/L) at five levels using clinical samples. A comparative evaluation between the two assays was performed using 187 clinical samples, and the concordance of qualitative results was also assessed. The repeatability and total imprecision (% coefficient of variation) of the Immulite TSI assay were 3.19% and 3.46% at 0.93 IU/L, and 3.76% and 5.42% at 19.3 IU/L, respectively. The linearity of this assay ranged from 0.16 to 6.17 IU/L. A high degree of correlation was observed between quantitative values from each assay (correlation coefficient = 0.819). Moderate agreement between methods was observed with an overall qualitative agreement of 93.0%. Among 13 cases with discordant qualitative results, the Immulite TSI assay generated more favorable results consistent with clinical diagnoses of patients than the Elecsys Anti-TSHR assay. The Immulite TSI assay showed reliable analytical performance and good correlation with the Elecsys Anti-TSHR assay and we expect this method will be helpful for clinicians to evaluate patients with hyperthyroidism.
Full Text
https://www.tandfonline.com/doi/full/10.1080/00365513.2019.1658216
DOI
10.1080/00365513.2019.1658216
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Dokyun(김도균) ORCID logo https://orcid.org/0000-0002-0348-5440
Kim, Byung Kwang(김병광)
Jeong, Seok Hoon(정석훈) ORCID logo https://orcid.org/0000-0001-9290-897X
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/173299
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links