Abstract
PURPOSE: The most common method of monitoring cardiac output (CO) is thermodilution using pulmonary artery catheter (PAC), but this method is associated with complications. Impedance cardiography (ICG) is a non-invasive CO monitoring technique. This study compared the accuracy and efficacy of ICG as a non-invasive cardiac function monitoring technique to those of thermodilution and arterial pressure contour.

MATERIALS AND METHODS: Sixteen patients undergoing liver transplantation were included. Cardiac index (CI) was measured by thermodilution using PAC, arterial waveform analysis, and ICG simultaneously in each patient. Statistical analysis was performed using intraclass correlation coefficient (ICC) and Bland-Altman analysis to assess the degree of agreement.

RESULTS: The difference by thermodilution and ICG was 1.13 L/min/m², and the limits of agreement were -0.93 and 3.20 L/min/m². The difference by thermodilution and arterial pressure contour was 0.62 L/min/m², and the limits of agreement were -1.43 and 2.67 L/min/m². The difference by arterial pressure contour and ICG was 0.50 L/min/m², and the limits of agreement were -1.32 and 2.32 L/min/m². All three percentage errors exceeded the 30% limit of acceptance. Substantial agreement was observed between CI of thermodilution with PAC and ICG at preanhepatic and anhepatic phases, as well as between CI of thermodilution and arterial waveform analysis at preanhepatic phase. Others showed moderate agreement.

CONCLUSION: Although neither method was clinically equivalent to thermodilution, ICG showed more substantial correlation with thermodilution method than with arterial waveform analysis. As a non-invasive cardiac function monitor, ICG would likely require further studies in other settings.
Files in This Item:
T201902818.pdf Download
DOI
10.3349/ymj.2019.60.8.735
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Anesthesiology and Pain Medicine (마취통증의학교실) > 1. Journal Papers
Yonsei Authors
Koo, Bon-Nyeo(구본녀) ORCID logo https://orcid.org/0000-0002-3189-1673
Kim, So Yeon(김소연) ORCID logo https://orcid.org/0000-0001-5352-157X
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/171094
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links