0 516

Cited 10 times in

A new low-profile visualized intraluminal support (LVIS) device, LVIS Blue: laboratory comparison between old and new LVIS

Authors
 Joonho Chung  ;  Yoshikazu Matsuda  ;  Jessica Nelson  ;  Kiffon Keigher  ;  Demetrius K Lopes 
Citation
 NEUROLOGICAL RESEARCH, Vol.40(1) : 78-85, 2018 
Journal Title
NEUROLOGICAL RESEARCH
ISSN
 0161-6412 
Issue Date
2018
Keywords
Intracranial aneurysm ; LVIS ; LVIS Blue ; in vitro experiments ; intracranial stent
Abstract
Objectives The Low-profile Visualized Intraluminal Support (LVIS) device is a flexible intracranial stent. The first generation of this system had significant challenges in consistently providing good wall apposition and aneurysm neck coverage. A new modified LVIS, LVIS Blue (Blue), has been developed to address these issues. The purpose of this study is to report a laboratory comparison of wall apposition and aneurysm neck coverage between the original LVIS and Blue. Methods In bench-top experiments, we noted the visual appearance of the devices and evaluated changes in stent cell angles, neck coverage surface area (%), and stent cell crossing profile for microcatheters using a fusiform aneurysm model. Our in vitro experiments included evaluation of wall apposition and aneurysm neck coverage of the devices under direct radiographic visualization. Results Blue showed three definite different zones (a mid-zone, a high-density zone, and a transitional zone) and higher metal coverage in the straight fusiform aneurysm model compared to LVIS. Two commercially available microcatheters easily crossed the stent cell at the greater curvature for both devices. In in vitro experiments, Blue showed better wall apposition in tortuous arteries and achieved higher neck coverage in the bifurcation aneurysm compared to LVIS. Discussion Blue achieved better wall apposition in tortuous arteries and higher aneurysm neck coverage (higher metal-to-artery ratio) in bifurcation aneurysms than LVIS. Our results may provide informative physical properties of LVIS and Blue to be expected when those are used for stent-assisted coil embolization of a large-giant fusiform aneurysm or a bifurcation saccular aneurysm clinically.
Full Text
https://www.tandfonline.com/doi/full/10.1080/01616412.2017.1398881
DOI
10.1080/01616412.2017.1398881
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Neurosurgery (신경외과학교실) > 1. Journal Papers