316 907

Cited 0 times in

근접치료용 방사성 동위원소의 선량분포 확인을 위한 디지털 반도체 센서의 제작 및 평가

Other Titles
 Evaluation of Fabricated Semiconductor Sensor for Verification of γ-ray Distribution in Brachytherapy 
Authors
 Jeong-Eun Park  ;  Kyo-Tae Kim  ;  Won-Hoon Choi  ;  Ho Lee  ;  Sam-Joo Cho  ;  So-Hyun Ahn  ;  Jin-Young Kim  ;  Yong-Keun Song  ;  Keum-bae Kim  ;  Hyun-Do Huh  ;  Sung-Kwang Park 
Citation
 Korean Journal of Medical Physics (의학물리), Vol.26(4) : 280-285, 2015 
Journal Title
Korean Journal of Medical Physics(의학물리)
ISSN
 1226-5829 
Issue Date
2015
Keywords
Digital radiation detector ; HgI2 ; γ-ray Detector ; Brachytherapy ; Radiation therapy
Abstract
In radiation therapy fields, a brachytherapy is a treatment that kills lesion of cells by inserting a radioisotope that keeps emitting radiation into the body. We currently verify the consistency of radiation treatment plan and dose distribution through film/screen system (F/S system), provide therapy after checking dose. When we check dose distribution, F/S systems have radiation signal distortion because there is low resolution by penumbra depending on the condition of film developed. In this study, We fabricated a HgI2 Semiconductor radiation sensor for base study in order that we verify the real dose distribution weather it's same as plans or not in brachytherapy. Also, we attempt to evaluate the feasibility of QA system by utilizing and evaluating the sensor to bracytherapy source. As shown in the result of detected signal with various source-to-detector distance (SDD), we quantitatively verified the real range of treatment which is also equivalent to treatment plans because only the low signal estimated as scatters was measured beyond the range of treatment. And the result of experiment that we access reproducibility on the same condition of γ-ray, we have made sure that the CV (coefficient of variation) is within 1.5 percent so we consider that the HgI2 sensor is available at QA of brachytherapy based on the result.
Files in This Item:
T201506787.pdf Download
DOI
10.14316/pmp.2015.26.4.280
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers
Yonsei Authors
Lee, Ho(이호) ORCID logo https://orcid.org/0000-0001-5773-6893
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/157393
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links