761 654

Cited 0 times in

전자기장 간섭이 자동제세동기에 미치는 영향: 실험연구 및 문헌고찰

DC Field Value Language
dc.contributor.author고동률-
dc.contributor.author공태영-
dc.contributor.author유제성-
dc.contributor.author정성필-
dc.contributor.author주영선-
dc.date.accessioned2017-10-26T07:34:34Z-
dc.date.available2017-10-26T07:34:34Z-
dc.date.issued2016-
dc.identifier.issn1226-4334-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/152212-
dc.description.abstractPurpose: Automated external defibrillators (AEDs) could not recommend shock for ventricular fibrillation in the presence of electromagnetic interference. The purpose of this study was to examine the effect of an induced electromagnetic field on performance of AED. Methods: The intensity of magnetic waves from commercial electric mats was measured. Three AEDs were attached to the resuscitation manikin and the question of whether shock would be recommended for simulated electrocardiogram of VF or normal sinus rhythm was tested. The simulation was repeated 10 times under the influence of 0, 5, and 18 μT magnetic field or electric mat. Relevant literature studies on electromagnetic interference on AED were reviewed. Results: The magnetic flux density from the electric mat was measured to 5.67-6.1 μT in warming phase, and 2.25-2.84 μT in maintenance phase. There was no false positive or false negative recommendation of shock under the influence of 0, 5, and 18 μT magnetic field or electric mat. However, one AED detected motion even in the stationary state. Among 11 studies from the literature search, five studies reported misinterpretation of AED. Minor errors including delayed analysis, motion artefact, and noise in speakers were reported from 6 studies. Conclusion: Although we could not reproduce false negative interpretation, AED made a mistake in confusing electromagnetic interference with motion artefact. Therefore, emergency providers should be cautious not to use AED close to household appliances or medical equipment inducing electromagnetic interference.-
dc.description.statementOfResponsibilityopen-
dc.formatapplication/pdf-
dc.languageKorean-
dc.publisher대한응급의학회-
dc.relation.isPartOfJournal of the Korean Society of Emergency Medicine (대한응급의학회지)-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.title전자기장 간섭이 자동제세동기에 미치는 영향: 실험연구 및 문헌고찰-
dc.title.alternativeEffect of Electromagnetic Interference on Automated External Defibrillator: Simulation Study with Literature Review.-
dc.typeArticle-
dc.publisher.locationKorea-
dc.contributor.collegeCollege of Medicine-
dc.contributor.departmentDept. of Emergency Medicine-
dc.contributor.googleauthor구재은-
dc.contributor.googleauthor유제성-
dc.contributor.googleauthor주영선-
dc.contributor.googleauthor공태영-
dc.contributor.googleauthor고동률-
dc.contributor.googleauthor정성필-
dc.contributor.localIdA00162-
dc.contributor.localIdA02507-
dc.contributor.localIdA03625-
dc.contributor.localIdA03955-
dc.contributor.localIdA00113-
dc.relation.journalcodeJ01868-
dc.contributor.alternativeNameKo, Dong Ryul-
dc.contributor.alternativeNameKong, Tae Young-
dc.contributor.alternativeNameYou, Je Sung-
dc.contributor.alternativeNameChung, Sung Pil-
dc.contributor.alternativeNameJoo, Young Seon-
dc.contributor.affiliatedAuthorKong, Tae Young-
dc.contributor.affiliatedAuthorYou, Je Sung-
dc.contributor.affiliatedAuthorChung, Sung Pil-
dc.contributor.affiliatedAuthorJoo, Young Seon-
dc.contributor.affiliatedAuthorKo, Dong Ryul-
dc.citation.volume27-
dc.citation.number3-
dc.citation.startPage231-
dc.citation.endPage237-
dc.identifier.bibliographicCitationJournal of the Korean Society of Emergency Medicine, Vol.27(3) : 231-237, 2016-
dc.date.modified2017-10-24-
dc.identifier.rimsid46989-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Emergency Medicine (응급의학교실) > 1. Journal Papers

qrcode