Evaluation of bilirubin interference and accuracy of six creatinine assays compared with isotope dilution-liquid chromatography mass spectrometry.

Authors
 Hyunjin Nah  ;  Sang-Guk Lee  ;  Kyeong-Seob Lee  ;  Jae-Hee Won  ;  Hyun Ok Kim  ;  Jeong-Ho Kim 
Citation
 CLINICAL BIOCHEMISTRY, Vol.49(3) : 274-281, 2016 
Journal Title
CLINICAL BIOCHEMISTRY
ISSN
 0009-9120 
Issue Date
2016
MeSH
Adolescent ; Adult ; Aged ; Aged, 80 and over ; Bilirubin/analysis* ; Bilirubin/blood ; Calibration ; Child ; Chromatography, Liquid/methods ; Creatinine/blood ; Female ; Glomerular Filtration Rate ; Humans ; Indicator Dilution Techniques ; Male ; Middle Aged ; Reference Standards ; Reproducibility of Results* ; Tandem Mass Spectrometry/methods
Keywords
Bilirubin interference ; Calibration traceability ; Creatinine ; ID–LC/MS
Abstract
OBJECTIVES: The aim of this study was to estimate bilirubin interference and accuracy of six routine methods for measuring creatinine compared with isotope dilution-liquid chromatography mass spectrometry (ID-LC/MS).
DESIGN AND METHODS: A total of 40 clinical serum samples from 31 patients with serum total bilirubin concentration >68.4μmol/L were collected. Serum creatinine was measured using two enzymatic reagents and four Jaffe reagents as well as ID-LC/MS. Correlations between bilirubin concentration and percent difference in creatinine compared with ID-LC/MS were analyzed to investigate bilirubin interference. Bias estimations between the six reagents and ID-LC/MS were performed. Recovery tests using National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 967a were also performed.
RESULTS: Both the enzymatic methods showed no bilirubin interference. However, three of the four Jaffe methods demonstrated significant bilirubin concentration-dependent interference in samples with creatinine levels <53μmol/L, and two of them showed significant bilirubin interference in samples with creatinine levels ranging from 53.0 to 97.2μmol/L. Comparison of these methods with ID-LC/MS using patients' samples with elevated bilirubin revealed that the tested methods failed to achieve the bias goal at especially low levels of creatinine. In addition, recovery test using NIST SRM 967a showed that bias in one Jaffe method and two enzymatic methods did not achieve the bias goal at either low or high level of creatinine, indicating they had calibration bias. One enzymatic method failed to achieve all the bias goals in both comparison experiment and recovery test.
CONCLUSIONS: It is important to understand that both bilirubin interference and calibration traceability to ID-LC/MS should be considered to improve the accuracy of creatinine measurement.
Full Text
http://www.sciencedirect.com/science/article/pii/S0009912015004816
DOI
10.1016/j.clinbiochem.2015.10.015
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Kim, Jeong Ho(김정호) ORCID logo https://orcid.org/0000-0003-2479-0548
Kim, Hyun Ok(김현옥) ORCID logo https://orcid.org/0000-0002-4964-1963
Nah, Hyunjin(나현진)
Lee, Sang-Guk(이상국) ORCID logo https://orcid.org/0000-0003-3862-3660
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/146348
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links