dc.description.abstract | We propose a simple isolated link circuit for a multi-channel bio-potential amplifier, by adapting the multiplexing/demultiplexing technique that is generally used in digital circuit design. The reduced number of isolation amplifiers, due to the multiplexing technique, results in an increase in the amount of isolation, as well as a reduction in manufacturing cost. The simple incorporation of a dead-zone in multiplexed timing, and a pull-down resistor in the multiplexed signal line, enables a multiple isolated bio-signal transmission over the shared isolation amplifier, with an affordable inter-channel cross talk. | - |
dc.description.statementOfResponsibility | open | - |
dc.format.extent | 469~476 | - |
dc.relation.isPartOf | PHYSIOLOGICAL MEASUREMENT | - |
dc.rights | CC BY-NC-ND 2.0 KR | - |
dc.rights.uri | https://creativecommons.org/licenses/by-nc-nd/2.0/kr/ | - |
dc.subject.MESH | Brain/physiology | - |
dc.subject.MESH | Electric Impedance | - |
dc.subject.MESH | Electroencephalography/instrumentation* | - |
dc.subject.MESH | Electroencephalography/methods* | - |
dc.subject.MESH | Equipment Design | - |
dc.subject.MESH | Humans | - |
dc.title | Improvement of multi-channel isolated bio-potential amplifiers using the multiplexing | - |
dc.type | Article | - |
dc.contributor.college | College of Medicine (의과대학) | - |
dc.contributor.department | Dept. of Medical Engineering (의학공학) | - |
dc.contributor.googleauthor | Sun K. Yoo | - |
dc.contributor.googleauthor | Nam H. Kim | - |
dc.contributor.googleauthor | Byung C. Chang | - |
dc.identifier.doi | 10.1088/0967-3334/23/2/320 | - |
dc.admin.author | false | - |
dc.admin.mapping | false | - |
dc.contributor.localId | A02471 | - |
dc.contributor.localId | A00357 | - |
dc.relation.journalcode | J02527 | - |
dc.identifier.eissn | 1361-6579 | - |
dc.identifier.pmid | 12051316 | - |
dc.identifier.url | http://iopscience.iop.org/0967-3334/23/2/320/ | - |
dc.contributor.alternativeName | Kim, Nam Hyun | - |
dc.contributor.alternativeName | Yoo, Sun Kook | - |
dc.contributor.affiliatedAuthor | Yoo, Sun Kook | - |
dc.contributor.affiliatedAuthor | Kim, Nam Hyun | - |
dc.rights.accessRights | not free | - |
dc.citation.volume | 23 | - |
dc.citation.number | 2 | - |
dc.citation.startPage | 469 | - |
dc.citation.endPage | 476 | - |
dc.identifier.bibliographicCitation | PHYSIOLOGICAL MEASUREMENT, Vol.23(2) : 469-476, 2002 | - |
dc.identifier.rimsid | 55575 | - |
dc.type.rims | ART | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.