0 499

Cited 17 times in

Comparative evaluation of two matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) systems, Vitek MS and Microflex LT, for the identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories

Authors
 Miae Lee  ;  Hae Sun Chung  ;  Hee Won Moon  ;  Sun Hwa Lee  ;  Kyungwon Lee 
Citation
 JOURNAL OF MICROBIOLOGICAL METHODS, Vol.113 : 13-15, 2015 
Journal Title
JOURNAL OF MICROBIOLOGICAL METHODS
ISSN
 0167-7012 
Issue Date
2015
MeSH
Gram-Positive Cocci/classification* ; Gram-Positive Cocci/isolation & purification* ; Humans ; Laboratories, Hospital ; Polymerase Chain Reaction ; Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization/methods*
Keywords
Bacterial species identification ; Gram-positive cocci ; Matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) ; Microflex LT ; Vitek MS
Abstract
We evaluated the performance of two MALDI-TOF MS systems for the identification of clinically important Gram-positive cocci. Vitek MS and Microflex LT correctly identified 97.2% and 94.7%, respectively. Both systems offer reliable and rapid identification of clinically important Gram-positive cocci isolated in clinical laboratories, including staphylococci, streptococci, and enterococci. Expanding the databases, especially of coagulase-negative staphylococci and viridans streptococci, would enhance performance.
Full Text
http://www.sciencedirect.com/science/article/pii/S0167701215000998
DOI
10.1016/j.mimet.2015.03.020
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Lee, Kyungwon(이경원) ORCID logo https://orcid.org/0000-0003-3788-2134
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/140253
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links