Comparison of prothrombin time derived from CoaguChek XS and laboratory test according to fibrinogen level

Authors
 Sue Jung Kim  ;  Eun Young Lee  ;  Rojin Park  ;  Juwon Kim  ;  Jaewoo Song 
Citation
 JOURNAL OF CLINICAL LABORATORY ANALYSIS, Vol.29(1) : 28-31, 2015 
Journal Title
JOURNAL OF CLINICAL LABORATORY ANALYSIS
ISSN
 0887-8013 
Issue Date
2015
MeSH
Analysis of Variance ; Blood Coagulation/drug effects ; Fibrinogen/metabolism* ; Humans ; International Normalized Ratio ; Point-of-Care Systems/standards* ; Prothrombin Time/methods* ; Reproducibility of Results
Keywords
CoaguChek XS ; INR ; fibrinogen ; point-of-care ; prothrombin time
Abstract
BACKGROUND: CoaguChek XS is one of the most widely used point-of-care (POC) devices to evaluate prothrombin time for monitoring oral anticoagulant therapy. Unlike laboratory methods, it detects electrical signals produced by thrombin activity to derive the international normalized ratio (INR). Therefore, we hypothesized that laboratory methods and CoaguChek XS could produce different results according to fibrinogen level.

METHODS: We compared INR values obtained from the CoaguChek XS and conventional laboratory method with 91 plasma samples covering a wide range of fibrinogen levels.

RESULTS: The samples were stratified into low, mid, and high fibrinogen groups by fibrinogen levels of <130 mg/dl, 130-450 mg/dl, and >450 mg/dl, respectively. The mean INR difference of the low fibrinogen group was significantly different from that of the mid or high fibrinogen group (P < 0.001). In the low fibrinogen group, CoaguChek XS INR showed a negative bias compared with the laboratory INR, while the mid and high fibrinogen groups had positive bias.

CONCLUSION: Our results suggest that patient selection according to fibrinogen status should precede the implementation of POC testing using CoaguChek XS. Also, periodic comparisons between CoaguChek XS and laboratory INR results should be continued during the use of CoaguChek XS.
Full Text
http://onlinelibrary.wiley.com/doi/10.1002/jcla.21722/abstract
DOI
10.1002/jcla.21722
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Laboratory Medicine (진단검사의학교실) > 1. Journal Papers
Yonsei Authors
Song, Jae Woo(송재우) ORCID logo https://orcid.org/0000-0002-1877-5731
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/139893
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links