364 528

Cited 0 times in

심박조율기 전극제거 중 발생한 심근파열

Other Titles
 Rupture of the Myocardium during the Removal of the Cardiac Pacemaker Electrode 
Authors
 이용경  ;  김지영  ;  곽영란  ;  송종욱  ;  오영준  ;  김승호 
Citation
 KOREAN JOURNAL OF ANESTHESIOLOGY, Vol.47(6) : 913-916, 2004 
Journal Title
KOREAN JOURNAL OF ANESTHESIOLOGY(대한마취과학회지)
ISSN
 2005-6419 
Issue Date
2004
Keywords
infection of pacemaker ; removal of pacemaker ; rupture of myocardium
Abstract
Infection of pacemaker electrode is one of the most frequent causes for the removal of the permanent cardiac pacemaker electrode and it is dangerous to remove the infected electrode by external traction method because the electrode is adhered to the myocardium. In this case, surgeon tried to remove infected cardiac pacing electrode using continuous external traction method. Myocardial rupture and consequent cardiac tamponade suddenly developed because the electrode was pulled by force. Emergent cardiopulmonary bypass was initiated and then, ruptured myocardium was repaired and the remaining electrode was removed without remarkable complications. This case emphasizes the risk of the myocardial rupture and the importance of preparing for emergent situation which can occur during the removal of the permanent pacemaker electrode, especially when the reason for the removal is the infection.
Files in This Item:
T200400347.pdf Download
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Anesthesiology and Pain Medicine (마취통증의학교실) > 1. Journal Papers
Yonsei Authors
Kwak, Young Lan(곽영란) ORCID logo https://orcid.org/0000-0002-2984-9927
Kim, Seung Ho(김승호)
Kim, Ji Young(김지영) ORCID logo https://orcid.org/0000-0001-5822-0338
Oh, Young Jun(오영준) ORCID logo https://orcid.org/0000-0002-6258-5695
Lee, Yong Kyung(이용경)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/112095
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links