1 592

Cited 84 times in

Problems associated with the 25-gauge transconjunctival sutureless vitrectomy system during and after surgery

DC Field Value Language
dc.contributor.author고형준-
dc.contributor.author권오웅-
dc.contributor.author김성수-
dc.contributor.author변석호-
dc.contributor.author이성철-
dc.date.accessioned2015-06-10T13:08:06Z-
dc.date.available2015-06-10T13:08:06Z-
dc.date.issued2006-
dc.identifier.issn0030-3755-
dc.identifier.urihttps://ir.ymlib.yonsei.ac.kr/handle/22282913/111072-
dc.description.abstractOBJECTIVE: To report the initial experiences with the 25-gauge transconjunctival sutureless vitrectomy (TSV) system, including intraoperative and postoperative problems. METHODS: We retrospectively reviewed the medical records and surgical videotapes of 50 consecutive patients who underwent vitrectomy performed by one surgeon using the TSV system. RESULTS: Intraoperatively, we encountered such problems as difficulty in inserting the microcannula, which led to deformity, instability of the microcannula, self-disconnection of the infusion tip and resultant lens damage, and conversion to 20-gauge conventional vitrectomy. Postoperatively, there were 8 cases with hypotony (IOP < 6 mm Hg) on day 1, 6 cases with elevated IOP, and 3 cases with retinal detachment during follow-up. CONCLUSION: Though certain problems exist during and after surgeries using TSV, this system is both convenient and safe for various vitreoretinal procedures.-
dc.description.statementOfResponsibilityopen-
dc.format.extent259~265-
dc.relation.isPartOfOPHTHALMOLOGICA-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.subject.MESHAdult-
dc.subject.MESHAged-
dc.subject.MESHEquipment Failure-
dc.subject.MESHFemale-
dc.subject.MESHFollow-Up Studies-
dc.subject.MESHHumans-
dc.subject.MESHIntraoperative Complications-
dc.subject.MESHLens, Crystalline/injuries-
dc.subject.MESHMale-
dc.subject.MESHMiddle Aged-
dc.subject.MESHOcular Hypotension/etiology*-
dc.subject.MESHPostoperative Complications-
dc.subject.MESHPrognosis-
dc.subject.MESHRetinal Detachment/surgery-
dc.subject.MESHRetrospective Studies-
dc.subject.MESHSuture Techniques/instrumentation*-
dc.subject.MESHVideo Recording-
dc.subject.MESHVitrectomy/adverse effects*-
dc.titleProblems associated with the 25-gauge transconjunctival sutureless vitrectomy system during and after surgery-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Ophthalmology (안과학)-
dc.contributor.googleauthorByeon S.H.-
dc.contributor.googleauthorChu Y.K.-
dc.contributor.googleauthorLee S.C.-
dc.contributor.googleauthorKoh H.J.-
dc.contributor.googleauthorKim S.S.-
dc.contributor.googleauthorKwon O.W.-
dc.identifier.doi10.1159/000093081-
dc.admin.authorfalse-
dc.admin.mappingfalse-
dc.contributor.localIdA00571-
dc.contributor.localIdA00152-
dc.contributor.localIdA00235-
dc.contributor.localIdA01849-
dc.contributor.localIdA02873-
dc.relation.journalcodeJ02431-
dc.identifier.eissn1423-0267-
dc.identifier.pmid16785758-
dc.identifier.urlhttp://www.karger.com/Article/FullText/93081-
dc.contributor.alternativeNameKoh, Hyoung Jun-
dc.contributor.alternativeNameKwon, Oh Woong-
dc.contributor.alternativeNameKim, Sung Soo-
dc.contributor.alternativeNameByeon, Suk Ho-
dc.contributor.alternativeNameLee, Sung Chul-
dc.contributor.affiliatedAuthorKim, Sung Soo-
dc.contributor.affiliatedAuthorKoh, Hyoung Jun-
dc.contributor.affiliatedAuthorKwon, Oh Woong-
dc.contributor.affiliatedAuthorByeon, Suk Ho-
dc.contributor.affiliatedAuthorLee, Sung Chul-
dc.rights.accessRightsnot free-
dc.citation.volume220-
dc.citation.number4-
dc.citation.startPage259-
dc.citation.endPage265-
dc.identifier.bibliographicCitationOPHTHALMOLOGICA, Vol.220(4) : 259-265, 2006-
dc.identifier.rimsid51960-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.