Periotest® Value와 Implant Stability Quotient에 영향을 미치는 요소

Other Titles
 THE INFLUENCING FACTORS OF PERIOTEST® VALUE AND IMPLANT STABILITY QUOTIENT 
Authors
 이영아  ;  차인호  ;  이호용  ;  한동후 
Citation
 Journal of Korean Academy of Prosthodontics (대한치과보철학회지), Vol.44(1) : 40-50, 2006 
Journal Title
Journal of Korean Academy of Prosthodontics(대한치과보철학회지)
ISSN
 0301-2875 
Issue Date
2006
Keywords
Periotest ; Osstell ; ISQ ; Implant stability
Abstract
Statement of problem: Periotest®and Osstell(TM) were known as the most objective and quantitative mobility tests available for evaluating stability of implant in vivo. Although a correlation between PTV widely used and ISQ recently introduced exist, a PTV was corresponded to various ISQ. A correct evaluation of implant stability could be obtained only after one has a thorough understanding of the limitations of devices and factors that affect measurements.

Purpose: The purpose of this study was to investigate the causes of variables in the values obtained with these two tests.

Material and method: A total of 333 implants 134 Branemark, 5 Silhouette and 194 ITI implants were investigated.

Result: 1. There was a correlation between PTV and ISQ (Spearman correlation =0.39, p<0.0001) 2. The factors that affected ISQ were diameter of implant future, location of implant and implant system (submerged type vs non-submerged type). 3. The factors that affected W were dimeter of implant future, location of implant, and elapsed time after implant placement. 4. There was no significant difference between different surface treatments of RBM, smooth surface and ti-unite on PTY and ISQ. 5. In radiographic finding, no saucerization or bone resorption has been detected in implants with ISQ values that were above the average level of each PTV. These higher values had higher bone densities around the implant fixture. Saucerization was observed in the most impants with ISQ values that were below the average level of each PTV.

Conclusion: There was a correlation between ISQ and PTV. However, each measuring methods had factors influencing the measured values. PTV were less sensitive to marginal bone resorption and influenced with the striking point on an implant to the level of bone. With ISQ, the height of implant from bone level to transducer should be considered.
Files in This Item:
T200600827.pdf Download
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Oral and Maxillofacial Surgery (구강악안면외과학교실) > 1. Journal Papers
2. College of Dentistry (치과대학) > Dept. of Prosthodontics (보철과학교실) > 1. Journal Papers
Yonsei Authors
Cha, In Ho(차인호) ORCID logo https://orcid.org/0000-0001-8259-2190
Han, Dong Hoo(한동후)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/109762
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links