204 483

Cited 0 times in

Use of a Tunneling Technique to Achieve a Lower Defibrillation Threshold during Implantable Cardioverter Defibrillator Implantation via the Right Subclavian Vein

Authors
 Jin-Bae Kim  ;  Boyoung Joung  ;  Moon-Hyoung Lee  ;  Sung-Soon Kim 
Citation
 JOURNAL OF KOREAN MEDICAL SCIENCE, Vol.25(10) : 1526-1528, 2010 
Journal Title
JOURNAL OF KOREAN MEDICAL SCIENCE
ISSN
 1011-8934 
Issue Date
2010
Keywords
Defibrillators, Implantable ; Electric Countershock
Abstract
A 56-yr-old man with aborted sudden cardiac death underwent implantable cardioverter defibrillator (ICD) implantation. While the ICD was being implanted, a left subclavian venogram failed to visualize the left subclavian vein, which was attributed to likely prolonged indwelling of the left subclavian sheath for venous access. Accordingly, the right subclavian vein was punctured and the ICD lead was diverted from the right side area to the active Can in the left pectoral area by tunneling over the sternum for high defibrillation threshold. The approach used in this case may be considered in patients who had difficult left subclavicular venous access and it may be prudent to save the left subclavian vein for ICD implantation in patients with fatal tachyarrhythmia.
Files in This Item:
T201005861.pdf Download
DOI
10.3346/jkms.2010.25.10.1526
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Sung Soon(김성순)
Lee, Moon-Hyoung(이문형) ORCID logo https://orcid.org/0000-0002-7268-0741
Joung, Bo Young(정보영) ORCID logo https://orcid.org/0000-0001-9036-7225
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/103226
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links