dc.description.abstractObjective: The aim of this study was to validate the Periotest values for the prediction of orthodontic mini-implants' stability. Methods: Sixty orthodontic mini-implants (7.0 mm x $\emptyset1.45$ mm; ACR, Biomaterials Korea, Seoul, Korea) were inserted into the buccal alveolar bone of 5 twelve month-old beagle dogs. Insertion torque (IT) and Periotest values (PTV) were measured at the installation procedure, and removal torque (RT) and PTV were recorded after 12 weeks of orthodontic loading. To correlate PTV with variables, the cortical bone thickness (mm) and bone mineral density (BMD) within the cortical bone and total bone area were calculated with the help of CT scanning. Results: The BMD and cortical bone thickness in mandibular alveolus were significantly higher than those of the maxilla (p < 0.05). The PTV values ranged from -3.2 to 4.8 for 12 weeks of loading showing clinically stable mini-implants. PTV at insertion was significantly correlated with IT (-0.51), bone density (-0.48), cortical bone thickness (-0.42) (p < 0.05) in the mandible, but showed no correlation in the maxilla. PTV before removal was significantly correlated with RT (-0.66) (p < 0.01) in the mandible. Conclusions: These results show that the periotest is a useful method for the evaluation of mini-implant stability, but it can only be applied to limited areas with thick cortical and high density bone such as the mandible.-
dc.description.statementOfResponsibilityopen-
dc.format.extent167~175-
dc.relation.isPartOfKOREAN JOURNAL OF ORTHODONTICS-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.title즉시 부하 교정용 미니임플랜트의 안정성 평가를 위한 Periotest ®의 유효성-
dc.title.alternativeThe validation of Periotest values for the evaluation of orthodontic mini-implants' stability-
dc.typeArticle-
dc.contributor.collegeCollege of Dentistry (치과대학)-
dc.contributor.departmentDept. of Orthodontics (교정과학)-
dc.contributor.googleauthor차정열-
dc.contributor.googleauthor유형석-
dc.contributor.googleauthor황충주-
dc.identifier.doi10.4041/kjod.2010.40.3.167-
dc.admin.authorfalse-
dc.admin.mappingfalse-
dc.contributor.localIdA02532-
dc.contributor.localIdA04006-
dc.contributor.localIdA04492-
dc.relation.journalcodeJ02087-
dc.identifier.eissn2005-372X-
dc.subject.keywordPeriotest-
dc.subject.keywordOrthodontic mini-implants-
dc.subject.keywordInsection torque-
dc.subject.keywordMobility-
dc.contributor.alternativeNameYu, Hyung Seog-
dc.contributor.alternativeNameCha, Jung Yul-
dc.contributor.alternativeNameHwang, Chung Ju-
dc.contributor.affiliatedAuthorYu, Hyung Seog-
dc.contributor.affiliatedAuthorCha, Jung Yul-
dc.contributor.affiliatedAuthorHwang, Chung Ju-
dc.citation.volume40-
dc.citation.number3-
dc.citation.startPage167-
dc.citation.endPage175-
dc.identifier.bibliographicCitationKOREAN JOURNAL OF ORTHODONTICS, Vol.40(3) : 167-175, 2010-
dc.identifier.rimsid51003-
dc.type.rimsART-
Appears in Collections:
2. College of Dentistry (치과대학) > Dept. of Orthodontics (교정과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.