비접촉 안압계, 리바운드 안압계, 토노펜, 골드만압평안압계의 안압 비교

Other Titles
 Comparison of Intraocular Pressure Measured by Non-Contact Tonometer, Rebound Tonometer, Tono-Pen, and Goldmann Applanation Tonometer 
Authors
 이경식  ;  김세경  ;  김응권  ;  김태임 
Citation
 JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, Vol.55(1) : 47-53, 2014 
Journal Title
JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY(대한안과학회지)
ISSN
 0378-6471 
Issue Date
2014
Keywords
Goldmann applanation tonometer ; Rebound tonometer
Abstract
Purpose : To compare the level of accuracy of intraocular pressure (IOP) measurements attained by non-contact tonometer (NCT), rebound tonometer (RT) Icare®, and Tono-Pen (TONO-PEN AVIA®), using Goldmann Applanation tonometer (GAT) as a reference value and to explore their clinical usefulness.

Methods : In a prospective study of 71 normal eyes, IOP was measured with NCT, RT, Tono-Pen and GAT. The IOP values of were then compared between the eyes.

Results : RT showed statistically most significant agreement with the GAT [ICC 0.811, 95%CI 0.712-0.878]. In analysis of Bland-Altman plots, NCT showed the smallest mean bias (+0.2 mm Hg) and widest CI (95%CI; ±5.05 mm Hg), RT showed relatively small mean bias (-0.7 mm Hg) and narrowest CI (95%CI; ±3.75 mm Hg).

Conclusions : There was a significant agreement between the RT and the GAT measurements. We expect RT to be considered as a reliable alternative when IOP measurement with GAT is not feasible.
Files in This Item:
T201403683.pdf Download
DOI
10.3341/jkos.2014.55.1.47
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Ophthalmology (안과학교실) > 1. Journal Papers
Yonsei Authors
Kim, Se Kyung(김세경)
Kim, Eung Kweon(김응권) ORCID logo https://orcid.org/0000-0002-1453-8042
Kim, Tae-Im(김태임) ORCID logo https://orcid.org/0000-0001-6414-3842
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/100134
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links