dc.description.abstractOBJECTIVES: The purpose of this study was to evaluate the most appropriate positioning of anatomical locking plates on the distal tibia. METHODS: We positioned 4 types of locking plates (3 straight and 1 bayonet) at anterior, middle, and posterior positions on the medial malleolus on both lower limbs of 15 cadavers. We measured the rotational mismatch angle between the tibia and the proximal part of the plate and calculated the plate prominence at each position. RESULTS: The average rotational mismatch angles of a combined group containing all 3 straight plates were 0.3, 16.9, and 29.4 degrees at the anterior, middle, and posterior positions, respectively (P < 0.001). In the group with 1 bayonet plate, the rotational mismatch angles were -10.5, 0, and 16.7 degrees on average, respectively (P < 0.001). The average plate prominence for each of the positions (anterior, middle, and posterior) in the group with 3 straight plates were 3.5, 6.9, and 9.4 mm, respectively (P < 0.001). The average plate prominence in the group with 1 bayonet plate were 5.4, 3.6, and 6.7 mm, respectively (P < 0.001). CONCLUSIONS: We confirmed that anterior placement of the 3 straight locking plates in the distal tibia showed the best fit with regard to rotational alignment and that middle placement of the bayonet locking plates showed best fit. Placement of distal end of the locking plates in inappropriate positions results in a greater risk of rotational mismatch and plate impingement.-
dc.description.statementOfResponsibilityopen-
dc.relation.isPartOfJOURNAL OF ORTHOPAEDIC TRAUMA-
dc.rightsCC BY-NC-ND 2.0 KR-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/2.0/kr/-
dc.subject.MESHAged-
dc.subject.MESHAged, 80 and over-
dc.subject.MESHAnkle Joint/physiology*-
dc.subject.MESHAnkle Joint/surgery*-
dc.subject.MESHBone Plates*-
dc.subject.MESHBone Screws*-
dc.subject.MESHCadaver-
dc.subject.MESHEquipment Design-
dc.subject.MESHEquipment Failure-
dc.subject.MESHEquipment Failure Analysis-
dc.subject.MESHFemale-
dc.subject.MESHHumans-
dc.subject.MESHInternal Fixators-
dc.subject.MESHMale-
dc.subject.MESHProsthesis Fitting/methods*-
dc.subject.MESHProsthesis Implantation/methods-
dc.subject.MESHRange of Motion, Articular-
dc.subject.MESHTibia/physiology*-
dc.subject.MESHTibia/surgery*-
dc.subject.MESHTreatment Outcome-
dc.titleAvoiding Rotational Mismatch of Locking Distal Tibia Plates Depends on Proper Plate Position-
dc.typeArticle-
dc.contributor.collegeCollege of Medicine (의과대학)-
dc.contributor.departmentDept. of Orthopedic Surgery (정형외과학)-
dc.contributor.googleauthorHyung-Keun Song-
dc.contributor.googleauthorJin-Wook Noh-
dc.contributor.googleauthorJae-Hoo Lee-
dc.contributor.googleauthorKyu-Hyun Yang-
dc.identifier.doi10.1097/BOT.0b013e31827cd67d-
dc.admin.authorfalse-
dc.admin.mappingfalse-
dc.contributor.localIdA02282-
dc.contributor.localIdA03090-
dc.relation.journalcodeJ01673-
dc.identifier.eissn1531-2291-
dc.identifier.pmid23249890-
dc.identifier.urlhttp://ovidsp.ovid.com/ovidweb.cgi?T=JS&CSC=Y&NEWS=N&PAGE=fulltext&AN=00005131-201307000-00013&LSLINK=80&D=ovft-
dc.subject.keyworddistal tibia-
dc.subject.keywordlocking plate-
dc.subject.keywordmalrotation-
dc.contributor.alternativeNameYang, Kyu Hyun-
dc.contributor.alternativeNameLee, Jae Hoo-
dc.contributor.affiliatedAuthorYang, Kyu Hyun-
dc.contributor.affiliatedAuthorLee, Jae Hoo-
dc.rights.accessRightsnot free-
dc.citation.volume27-
dc.citation.number7-
dc.citation.startPage147-
dc.citation.endPage151-
dc.identifier.bibliographicCitationJOURNAL OF ORTHOPAEDIC TRAUMA, Vol.27(7) : 147-151, 2013-
dc.identifier.rimsid32970-
dc.type.rimsART-
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Orthopedic Surgery (정형외과학교실) > 1. Journal Papers

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.