1 2

Cited 0 times in

Cited 0 times in

Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems

Authors
 Oh, Ji Yoon  ;  Kim, You Rim  ;  Lee, Yong Ju  ;  Jeon, Dae Keun  ;  Choi, Ji Soo  ;  Uhm, Jae-Sun  ;  Ahn, Chul-Min  ;  Jang, Won Seuk 
Citation
 MEDICAL SCIENCE MONITOR, Vol.31, 2025-12 
Article Number
 e949932 
Journal Title
MEDICAL SCIENCE MONITOR
ISSN
 1234-1010 
Issue Date
2025-12
MeSH
Adult ; Aged ; Algorithms ; Arrhythmias, Cardiac* / diagnosis ; Arrhythmias, Cardiac* / physiopathology ; Clinical Alarms* ; Detection Algorithms ; Electrocardiography / instrumentation ; Electrocardiography / methods ; Female ; Humans ; Intensive Care Units ; Male ; Middle Aged ; Monitoring, Physiologic / instrumentation ; Monitoring, Physiologic / methods ; Workload
Keywords
Arrhythmias, Cardiac ; Electrocardiography ; Intensive Care Units ; Monitoring, Physiologic
Abstract
Background: Patient monitoring systems are widely used in intensive care units (ICUs) to monitor patient's conditions. A high false alarm rate can lead to alarm fatigue among nurses, increasing workload and stress. This study aimed to improve the accuracy of arrhythmia detection by enhancing the noise detection algorithm in patient monitoring systems and to determine whether false alarm rate and workload decreased through clinical trials. Material/Methods: Trials were conducted on adult patients in the ICU at Yongin Severance Hospital who required continuous electrocardiogram (ECG) monitoring for at least 2 days. After the first trial, the noise detection algorithm of the M50 (investigational device) was improved, and a second trial was conducted to evaluate its performance. Both trials followed the same study design. During the study period, M50 and MX700 (comparator device) were applied simultaneously for 3 days. Arrhythmia alarms were reviewed by an independent evaluator who assessed false alarms by comparing them with the ECG signals. False alarm rates were compared between trials using the chi-square (c2) test. Results: The clinical trial was conducted through 2 separate trials, with 17 and 11 participants, respectively. A comparative analysis of false alarm rates of the investigational device demonstrated a reduction from 71.75% to 27.61%. Statistical analysis using the chi-square test indicated a P value of 0.000 (<0.001), confirming a statistically significant difference. Conclusions: The results of 2 trials demonstrated reductions in false alarm rate and NASA-TLX score. These findings suggest that enhancing the noise detection algorithm in the patient monitoring system improved arrhythmia detection accuracy and helped reduce nurses' workload.
Files in This Item:
90878.pdf Download
DOI
10.12659/MSM.949932
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Internal Medicine (내과학교실) > 1. Journal Papers
1. College of Medicine (의과대학) > BioMedical Science Institute (의생명과학부) > 1. Journal Papers
Yonsei Authors
Ahn, Chul-Min(안철민)
Uhm, Jae Sun(엄재선) ORCID logo https://orcid.org/0000-0002-1611-8172
Chang, Won Seok(장원석) ORCID logo https://orcid.org/0000-0002-4925-3549
Choi, Ji Soo(최지수) ORCID logo https://orcid.org/0000-0003-2716-6775
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/210040
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links