Feasibility study of patient-specific four-dimensional in vivo tracking system for high-dose-rate brachytherapy: Experimental evaluation

Authors
 Oh, Geon  ;  Lee, Jeongshim  ;  Kim, Hunjung  ;  Cheon, Wonjoong  ;  Shin, Dong-seok  ;  Seo, Jaehyeon  ;  Sung, Jiwon  ;  Shin, Dongho  ;  Yoon, Myonggeun  ;  Chung, Jin-beom  ;  Lee, Boram 
Citation
 MEDICAL PHYSICS, Vol.52(4) : 2533-2550, 2025-04 
Journal Title
MEDICAL PHYSICS
ISSN
 0094-2405 
Issue Date
2025-04
MeSH
Brachytherapy* / instrumentation ; Brachytherapy* / methods ; Feasibility Studies ; Humans ; Iridium Radioisotopes / therapeutic use ; Monte Carlo Method ; Phantoms, Imaging ; Radiation Dosage* ; Radiotherapy Dosage
Keywords
HDR brachytherapy ; in vivo 4D tracking system ; patient-specific method
Abstract
BackgroundHigh-dose-rate (HDR) brachytherapy using Iridium-192 as a radiation source is widely employed in cancer treatment to deliver concentrated radiation doses while minimizing normal tissue exposure. In this treatment, the precision with which the sealed radioisotope source is delivered significantly impacts clinical outcomes.PurposeThis study aims to evaluate the feasibility of a new four-dimensional (4D) in vivo source tracking and treatment verification system for HDR brachytherapy using a patient-specific approach.MethodsA hardware system was developed for the experiments, featuring a high-resolution compact gamma camera with a redesigned diverging collimator, enhanced detector, and precision control system. The collimator was redesigned to improve spatial resolution by reducing the hole size and increasing the hole array, while reducing the pixel size of the detector and increasing the number of pixels. The performance was evaluated using Monte Carlo simulations, which demonstrated significant improvements in spatial resolution. Experiments were conducted in a controlled setup using a phantom to simulate clinical conditions. The phantom was positioned at various distances from the gamma camera (327.30, 377.30, and 427.30 mm) and imaged at multiple angles. The accuracy of the system was tested in four different cases: three with fixed distances and one employing a multi-focusing method. The multi-focusing method allows the gamma camera to adjust its focus based on the anatomical characteristics of individual patients, thereby enhancing source-tracking accuracy. The performance of the system was evaluated under these four different scenarios. The Euclidean distance and three-dimensional gamma analysis were used to evaluate tracking accuracy and dose distribution.ResultsThe redesigned collimator demonstrated significant improvements (compared to the previous design) in the spatial resolution of the gamma camera, showing 34.21% and 23.46% enhancements in the horizontal and vertical profiles, respectively. These improvements in gamma camera resolution are crucial for enhancing the tracking system's accuracy. The experimental results demonstrated varying degrees of accuracy across different cases, reflecting the performance of the system under different conditions. The average Euclidean distance errors were Case 1 (327.30 mm): 1.358 mm; Case 2 (377.30 mm): 1.731 mm; Case 3 (427.30 mm): 1.973 mm; and Case 4 (multi-focusing): 1.527 mm. The gamma pass rates for the four cases were:- Case 1: 86.39%; Case 2: 75.28%; Case 3: 72.22%; and Case 4: 81.67% (1 mm/1%). For the 2 mm/2% criterion, the gamma pass rates were 97.11, 94.72, 92.38, and 96.78% for Cases 1, 2, 3, and 4, respectively. Case 4 (multi-focusing) showed an improvement over Case 3, with a 22.6% reduction in the average Euclidean distance error and a 13.1% increase in the gamma pass rate (1 mm/1%).ConclusionThese results demonstrate that the new 4D in vivo source tracking and treatment verification system for HDR brachytherapy is feasible and has potential clinical benefits.
Full Text
https://aapm.onlinelibrary.wiley.com/doi/10.1002/mp.17614
DOI
10.1002/mp.17614
Appears in Collections:
1. College of Medicine (의과대학) > Dept. of Radiation Oncology (방사선종양학교실) > 1. Journal Papers
Yonsei Authors
Sung, Jiwon(성지원)
URI
https://ir.ymlib.yonsei.ac.kr/handle/22282913/208877
사서에게 알리기
  feedback

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse

Links